共 50 条
Biomechanical evaluation of atlantoaxial transarticular screw fixation technique
被引:0
|作者:
瞿东滨
金大地
赵卫东
钟世镇
机构:
关键词:
D O I:
暂无
中图分类号:
R615 [手术基本操作技术];
学科分类号:
摘要:
Objective: To compare the biomechanical differences among the atlantoaxial transarticular screw fixation (Magerl) and other posterior fixation techniques. Methods: Seven preserved atlantoaxial complex specimens were harvested and fixated with Magerl, Magerl plus Gallie wiring, Magerl plus Brooks wiring, pure Brooks, pure Gallie, and Halifax interlaminar clamping fixation, respectively. The torque for every fixation technique was measured at the point of five degree rotation of the atlantoaxial joint. Results: The torque for Magerl was 6.59 N·m± 1.14 N·m , which was significantly higher than any other pure posterior techniques including Gallie ( 1.74 N·m± 0.31 N·m ). Brooks ( 4.06 N·m± 0.48 N·m) and Halifax ( 3.44 N·m± 0.87 N·m) (P< 0.01 ), but less than Magerl plus brooks ( 9.94 N·m± 1.45 N·m) (P< 0.01 ). No statistically significant difference was found between Magerl and Magerl plus Gallie wiring ( 7.61 N·m± 1.10 N·m) or between Brooks and Halifax. Conclusions: Compared with other pure posterior fixation techniques the atlantoaxial transarticular screw fixation technique provides more torsion resistance capacity. It is also suggested that combined Gallie wiring do not add any biomechanical superiority to this technique.
引用
收藏
页数:3
相关论文