Reproducibility of Quantitative Double-Echo Steady-State T2 Mapping of Knee Cartilage

被引:1
|
作者
Williams, Ashley A. [2 ,3 ]
Asay, Jessica L. [4 ]
Asare, Daniella [3 ]
Desai, Arjun D. [4 ,5 ]
Gold, Garry E. [4 ,6 ]
Hargreaves, Brian A. [4 ,5 ,6 ]
Chaudhari, Akshay S. [4 ,7 ]
Chu, Constance R. [1 ,2 ,3 ]
机构
[1] Stanford Orthoped, Orthoped Surg, 450 Broadway St, Redwood City, CA 94063 USA
[2] Stanford Univ, Dept Orthopaed Surg, Stanford, CA USA
[3] VA Palo Alto Hlth Care Syst, Palo Alto, CA USA
[4] Stanford Univ, Dept Radiol, Stanford, CA USA
[5] Stanford Univ, Dept Elect Engn, Stanford, CA USA
[6] Stanford Univ, Dept Bioengn, Stanford, CA USA
[7] Stanford Univ, Dept Biomed Data Sci, Stanford, CA USA
关键词
T-2; mapping; cartilage; reproducibility; DESS; knee; T2; RELAXATION-TIMES; JOINT CARTILAGE; OSTEOARTHRITIS; MRI; HEALTHY; RELIABILITY; MORPHOMETRY; T-1-RHO;
D O I
10.1002/jmri.29431
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Cartilage T-2 can detect joints at risk of developing osteoarthritis. The quantitative double-echo steady state (qDESS) sequence is attractive for knee cartilage T-2 mapping because of its acquisition time of under 5 minutes. Understanding the reproducibility errors associated with qDESS T-2 is essential to profiling the technical performance of this biomarker. Purpose: To examine the combined acquisition and segmentation reproducibility of knee cartilage qDESS T-2 using two different regional analysis schemes: 1) manual segmentation of subregions loaded during common activities and 2) automatic subregional segmentation. Study Type: Prospective. Subjects: 11 uninjured participants (age: 28 +/- 3 years; 8 (73%) female). Field Strength/Sequence: 3-T, qDESS. Assessment: Test-retest T-2 maps were acquired twice on the same day and with a 1-week interval between scans. For each acquisition, average cartilage T-2 was calculated in four manually segmented regions encompassing tibiofemoral contact areas during common activities and 12 automatically segmented regions from the deep-learning open-source framework for musculoskeletal MRI analysis (DOSMA) encompassing medial and lateral anterior, central, and posterior tibiofemoral regions. Test-retest T-2 values from matching regions were used to evaluate reproducibility. Statistical Tests: Coefficients of variation (%CV), root-mean-square-average-CV (%RMSA-CV), and intraclass correlation coefficients (ICCs) assessed test-retest T-2 reproducibility. The median of test-retest standard deviations was used for T-2 precision. Bland-Altman (BA) analyses examined test-retest biases. The smallest detectable difference (SDD) was defined as the BA limit of agreement of largest magnitude. Significance was accepted for P < 0.05. Results: All cartilage regions across both segmentation schemes demonstrated intraday and interday qDESS T-2 CVs and RMSA-CVs of <= 5%. T-2 ICC values >0.75 were observed in the majority of regions but were more variable in interday tibial comparisons. Test-retest T-2 precision was <1.3 msec. The T-2 SDD was 3.8 msec. Data Conclusion: Excellent CV and RMSA-CV reproducibility may suggest that qDESS T-2 increases or decreases >5% (3.8 msec) could represent changes to cartilage composition.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Editorial for "Reproducibility of Quantitative Double-Echo Steady-State T2 Mapping of Knee Cartilage"
    Bugeja, Jessica M.
    Pagnozzi, Alex M.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2024,
  • [2] Imaging and T2 relaxometry of short-T2 connective tissues in the knee using ultrashort echo-time double-echo steady-state (UTEDESS)
    Chaudhari, Akshay S.
    Sveinsson, Bragi
    Moran, Catherine J.
    McWalter, Emily J.
    Johnson, Ethan M.
    Zhang, Tao
    Gold, Garry E.
    Hargreaves, Brian A.
    MAGNETIC RESONANCE IN MEDICINE, 2017, 78 (06) : 2136 - 2148
  • [3] Five-minute knee MRI for simultaneous morphometry and T2 relaxometry of cartilage and meniscus and for semiquantitative radiological assessment using double-echo in steady-state at 3T
    Chaudhari, Akshay S.
    Black, Marianne S.
    Eijgenraam, Susanne
    Wirth, Wolfgang
    Maschek, Susanne
    Sveinsson, Bragi
    Eckstein, Felix
    Oei, Edwin H. G.
    Gold, Garry E.
    Hargreaves, Brian A.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2018, 47 (05) : 1328 - 1341
  • [4] Quantitative double echo steady state T2 mapping of upper extremity peripheral nerves and muscles
    Campbell, Gracyn J.
    Sneag, Darryl B.
    Queler, Sophie C.
    Lin, Yenpo
    Li, Qian
    Tan, Ek T.
    FRONTIERS IN NEUROLOGY, 2024, 15
  • [5] T2 Mapping of Cartilage with Triple-Echo Steady State MR Sequence
    Juras, V.
    Bohndorf, K.
    Heule, R.
    Kronnerwetter, C.
    Szomolanyi, P.
    Bieri, O.
    Frollo, I.
    Trattnig, S.
    2017 11TH INTERNATIONAL CONFERENCE ON MEASUREMENT, 2017, : 189 - 192
  • [6] Characterization of the collagen component of cartilage repair tissue of the talus with quantitative MRI: comparison of T2 relaxation time measurements with a diffusion-weighted double-echo steady-state sequence (dwDESS)
    M. Kretzschmar
    O. Bieri
    M. Miska
    M. Wiewiorski
    N. Hainc
    V. Valderrabano
    U. Studler
    European Radiology, 2015, 25 : 980 - 986
  • [7] Characterization of the collagen component of cartilage repair tissue of the talus with quantitative MRI: comparison of T2 relaxation time measurements with a diffusion-weighted double-echo steady-state sequence (dwDESS)
    Kretzschmar, M.
    Bieri, O.
    Miska, M.
    Wiewiorski, M.
    Hainc, N.
    Valderrabano, V.
    Studler, U.
    EUROPEAN RADIOLOGY, 2015, 25 (04) : 980 - 986
  • [8] Comparison of three-dimensional double-echo steady-state sequence with routine two-dimensional sequence in the depiction of knee cartilage
    Marangaloo S.L.
    Ariamanesh A.S.
    Aminzadeh B.
    Abedi H.
    Abbaszadeh A.
    Montazerabadi A.
    Iranian Journal of Medical Physics, 2020, 17 (02) : 75 - 80
  • [9] A comparison of multi-echo spin-echo and triple-echo steady-state T2 mapping for in vivo evaluation of articular cartilage
    Vladimir Juras
    Klaus Bohndorf
    Rahel Heule
    Claudia Kronnerwetter
    Pavol Szomolanyi
    Benedikt Hager
    Oliver Bieri
    Stefan Zbyn
    Siegfried Trattnig
    European Radiology, 2016, 26 : 1905 - 1912
  • [10] A comparison of multi-echo spin-echo and triple-echo steady-state T2 mapping for in vivo evaluation of articular cartilage
    Juras, Vladimir
    Bohndorf, Klaus
    Heule, Rahel
    Kronnerwetter, Claudia
    Szomolanyi, Pavol
    Hager, Benedikt
    Bieri, Oliver
    Zbyn, Stefan
    Trattnig, Siegfried
    EUROPEAN RADIOLOGY, 2016, 26 (06) : 1905 - 1912