Bilateral comparison of 1.018 V and 10 V standards between DEFNAT (Tunisia) and the BIPM, April to June 2022 (part of the ongoing BIPM key comparison BIPM.EM-K11.a and b)

被引:1
|
作者
Solve S. [1 ]
Chayramy R. [1 ]
Stock M. [1 ]
Zouaoui J. [2 ]
机构
[1] Bipm, Pavillon de Breteuil Cedex, Sèvres
[2] Defnat, Laboratoire de Metrologie Electrique Direction Generale des Transmissions et de l'Informatique, Base Militaire Bab Saâdoun 1068 Cité Rommana BP, Tunis
关键词
D O I
10.1088/0026-1394/60/1A/01003
中图分类号
学科分类号
摘要
Main text As part of the ongoing BIPM key comparison BIPM.EM-K11.a and b, a comparison of the 1.018 V and 10 V voltage reference standards of the BIPM and the Laboratoire de Métrologie Électrique, DEFNAT (Tunisia), was carried out from April to June 2022. Two BIPM Zener diode-based travelling standards (Fluke 732B), BIPM_E (ZE) and BIPM_F (ZF), were transported by freight to DEFNAT to be measured and back to the BIPM for checking the stability of the traveling standards. At DEFNAT, the reference standard for DC voltage is a Josephson Voltage Standard (JVS). The output EMF (Electromotive Force) of each travelling standard was measured by direct comparison with the primary standard. The final result of the comparison is presented as the difference between the values assigned to DC voltage standards by DEFNAT, at the level of 1.018 V and 10 V, at DEFNAT, U DEFNAT, and those assigned by the BIPM, at the BIPM, U BIPM, at the reference date of the 15th of May 2022. U DEFNAT - U BIPM = 0.03 μV; u c = 0.03 μV, at 1.018 V U DEFNAT - U BIPM = 0.44 μV; u c = 0.12 μV, at 10 V where u c is the combined standard uncertainty associated with the measured difference, including the uncertainty of the representation of the volt at the BIPM and at DEFNAT, based on K J, and the uncertainty related to the comparison. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA). © 2023 BIPM & IOP Publishing Ltd.
引用
收藏
相关论文
共 50 条
  • [31] Comparison of the Josephson Voltage Standards of the SMD and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b)
    Solve, S.
    Chayramy, R.
    Stock, M.
    Holiastou, M.
    Flouda, I.
    [J]. METROLOGIA, 2010, 47
  • [32] Comparison of the Josephson Voltage Standards of the DMDM and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b)
    Solve, S.
    Chayramy, R.
    Stock, M.
    Pantelic-Babic, J.
    Sofranac, Z.
    Vujovic, T. Cincar
    [J]. METROLOGIA, 2016, 53
  • [33] Comparison of the josephson voltage standards of the NIM and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b)
    Solve, S.
    Chayramy, R.
    Stock, M.
    Yuan, Gao
    Honghui, Li
    Zengmin, Wang
    [J]. Metrologia, 2014, 51 (1A)
  • [34] Comparison of the Josephson Voltage Standards of the CMI and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b)
    Solve, S.
    Chayramy, R.
    Stock, M.
    Streit, Jiri
    Sira, Martin
    [J]. METROLOGIA, 2012, 49
  • [35] Comparison of the Josephson Voltage Standards of the INM and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b)
    Solve, S.
    Chayramy, R.
    Stock, M.
    Simionescu, M.
    Cirneanu, L.
    [J]. METROLOGIA, 2015, 52
  • [36] Comparison of the Josephson Voltage Standards of the PTB and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b)
    Solve, S.
    Chayramy, R.
    Stock, M.
    Palafox, L.
    Behr, R.
    [J]. METROLOGIA, 2015, 52
  • [37] Bilateral comparison of 10 kΩ standards (ongoing BIPM key comparison BIPM.EM-K13.b) between the CMI (Czech Republic) and the BIPM
    Goebel, R.
    Chrobok, P.
    Fletcher, N.
    Stock, M.
    [J]. METROLOGIA, 2008, 45
  • [38] Comparison of the Josephson Voltage Standards of the INTI and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.a)
    Solve, S.
    Chayramy, R.
    Stock, M.
    Iuzzolino, R.
    Tonina, A.
    [J]. METROLOGIA, 2012, 49
  • [39] Bilateral comparison of 1 ω and 10 kω standards (ongoing BIPM key comparison BIPM.EM-K13) between the NPLI (India) and the BIPM
    Rolland B.
    Stock M.
    Gournay P.
    Aloysius R.P.
    Kiran U.
    Hemavathi A.
    Awana V.P.S.
    Biswas J.C.
    [J]. Metrologia, 2022, 59 (1 A)
  • [40] Bilateral comparison of 10 V standards between the NSAI - NML (Ireland) and the BIPM (BIPM.EM-K11.b)
    Solve, Stephane
    Chayramy, Regis
    Stock, Michael
    Power, Oliver