Radiofrequency ablation for the treatment of great saphenous venous insufficiency using a newly developed VENISTAR catheter: A preliminary study

被引:0
|
作者
Min, Jeeyoung [1 ]
Hwang, Jin Ho [1 ,3 ]
Park, Sang Woo [1 ]
Hwang, Jae Joon [2 ]
Kim, Jun Seok [2 ]
Lee, Song Am [2 ]
Chee, Hyun Keun [2 ]
机构
[1] Konkuk Univ, Dept Radiol, Sch Med, Seoul, South Korea
[2] Konkuk Univ, Sch Med, Dept Thorac & Cardiovasc Surg, Seoul, South Korea
[3] Konkuk Univ, Dept Radiol, Sch Med, 120 Neungdong Ro, Seoul 05029, South Korea
关键词
Radiofrequency ablation; new catheter; chronic venous insufficiency; great saphenous vein; CLINICAL-PRACTICE-GUIDELINES; VARICOSE-VEINS; VASCULAR-SURGERY; LASER-ABLATION; MULTICENTER; SOCIETY;
D O I
10.1177/17085381241250112
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objectives This study was aimed to assess the preliminary outcomes of radiofrequency ablation (RFA) using a newly developed catheter (VENISTAR) for the treatment of incompetent great saphenous veins (GSVs). Methods In this prospective observational study, endovenous RFA using a VENISTAR catheter was performed on 16 saphenous veins in 12 patients between August and November 2019. Patients' pre- and post-procedural data were recorded. Doppler ultrasound imaging and clinical evaluation were performed at 1 week and 1, 3, and 6 months to determine the efficacy and safety of the treatment. Results Technical success and complete closure of the targeted GSVs immediately after the procedure were observed in all 16 limbs (100%). However, one patient (one limb) was found to have partial occlusion without significant reflux after 1 week of follow-up. Kaplan-Meier analysis yielded a complete occlusion rate of 93% at 6 months of follow-up. The Venous Clinical Severity Scores at the time of all follow-up were significantly lower than those at baseline (3.3 +/- 1.1 at baseline to 0.6 +/- 0.6, 0.3 +/- 0.6, 0.1 +/- 0.4, and 0.2 +/- 0.4 at 1 week and 1, 3, and 6 months, respectively) (p < .001). Mild post-procedural pain was noted in 7 and 4 limbs at 1 week and 1 month, respectively. Grade 1 ecchymosis over the ablated segment was noted in 5 (35.7%) of 14 limbs at 1-week follow-up. Conclusions Endovenous treatment of GSV insufficiency using a new VENISTAR radiofrequency catheter has been shown to be feasible, effective, and safe throughout the 6-month follow-up.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Midterm Outcome of Mechanochemical Endovenous Ablation for the Treatment of Great Saphenous Vein Insufficiency
    Witte, Marianne E.
    Holewijn, Suzanne
    van Eekeren, Ramon R.
    de Vries, Jean-Paul
    Zeebregts, Clark J.
    Reijnen, Michel M. P. J.
    JOURNAL OF ENDOVASCULAR THERAPY, 2017, 24 (01) : 149 - 155
  • [22] Great saphenous vein sparing techniques in lower extremity venous insufficiency treatment
    Bademci, Mehmet Senel
    Kocaaslan, Cemal
    Bayraktar, Fatih Avni
    Oztekin, Ahmet
    Aydin, Bilal
    Aydin, Ebuzer
    PHLEBOLOGY, 2021, 36 (04) : 331 - 331
  • [23] A comparative study of 1,470-nm endovenous laser ablation and segmental radiofrequency ablation in the treatment of saphenous veins insufficiency
    Torma, N.
    Frankovicova, M.
    BRATISLAVA MEDICAL JOURNAL-BRATISLAVSKE LEKARSKE LISTY, 2021, 122 (12): : 876 - 879
  • [24] Outcomes of Cyanoacrylate Closure Versus Radiofrequency Ablation for the Treatment of Incompetent Great Saphenous Veins
    Alhewy, Mohammed Alsagheer
    Abdo, Ehab M.
    Ghazala, Ehab Abd elmoneim
    Khamis, Ahmed Atef
    Gado, Hassan
    Abd-Elgawad, Wael Abdo Abdo
    Abdelhafez, Abdelaziz Ahmed
    El Sayed, Abdullah
    Khedr, Alhussein M.
    Mosaed, Haytham Ameer Mahmoud
    ANNALS OF VASCULAR SURGERY, 2024, 98 : 309 - 316
  • [25] ENDOVENOUS ABLATION WITH LASER FOR THE TREATMENT OF THE GREAT SAPHENOUS VEIN INSUFFICIENCY. IS IT SAFE AND EFFECTIVE?
    Memetoglu, M. E.
    Kalkan, A.
    Kurtcan, S.
    Kara, V.
    Ertas, G.
    Cetinkaya, A. S.
    Talay, S.
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2011, 147 : S93 - S93
  • [26] Our mid-and long-term results of endovenous radiofrequency ablation in great saphenous vein insufficiency
    Oguz, Sonay
    ANNALS OF CLINICAL AND ANALYTICAL MEDICINE, 2024, 15 (07): : 519 - 522
  • [27] A randomized controlled noninferiority trial comparing radiofrequency with stripping and conservative hemodynamic cure for venous insufficiency technique for insufficiency of the great saphenous vein
    Gonzalez Canas, Elena
    Florit Lopez, Salvador
    Vives Vilagut, Roser
    Alejandro Guevara-Noriega, Kerbi
    Santos Espi, Marta
    Rios, Jose
    Navarro Soto, Salvador
    Gimenez Gaibar, Antonio
    JOURNAL OF VASCULAR SURGERY-VENOUS AND LYMPHATIC DISORDERS, 2021, 9 (01) : 101 - 112
  • [28] Endovenous Treatment for Great Saphenous Vein Insufficiency: A Comparative Study of Segmental Radiofrequency and 1470-nm Endovenous Laser
    Filippo, Marcio Gomes
    Harduin, Leonardo de Oliveira
    Barroso, Thiago Almeida
    de Almeida, Leonardo Cortizo
    Schanaider, Alberto
    Espinosa, Gaudencio
    VASCULAR AND ENDOVASCULAR SURGERY, 2025,
  • [29] Single-center retrospective review of early outcomes of radiofrequency ablation versus cyanoacrylate ablation of isolated great saphenous vein insufficiency
    Bademci, Mehmet Senel
    Kocaaslan, Cemal
    Aldag, Mustafa
    Yalvac, Emine Seyma Denli
    Oztekin, Ahmet
    Bulut, Nurgul
    Aydin, Ebuzer
    JOURNAL OF VASCULAR SURGERY-VENOUS AND LYMPHATIC DISORDERS, 2019, 7 (04) : 480 - 485
  • [30] Comparative efficacy and life quality effects of surgical stripping, radiofrequency ablation, and cyanoacrylate embolization in patients undergoing treatment for great saphenous vein insufficiency
    Ay, Yasin
    Gunes, Esra
    Turkkolu, Sevket Tuna
    Selcuk, Emre
    Calim, Muhittin
    Akal, Ramazan
    Aydin, Cemalettin
    Inan, Bekir
    Koksal, Cengiz
    Ay, Nuray Kahraman
    PHLEBOLOGY, 2021, 36 (01) : 54 - 62