Brightness evaluation of pulsed electron gun using negative electron affinity photocathode developed for time-resolved measurement using scanning electron microscope

被引:7
|
作者
Morishita H. [1 ,2 ]
Ohshima T. [1 ]
Otsuga K. [1 ]
Kuwahara M. [3 ]
Agemura T. [4 ]
Ose Y. [4 ]
机构
[1] Research & Development Group, Hitachi, Ltd., 1-280, Higashi-koigakubo Kokubunji-shi, Tokyo
[2] Graduate School of Engineering, Nagoya University, Furo-cho, Chikusa, Nagoya-shi, Aichi
[3] Institute of Materials and Systems for Sustainability, Nagoya University, Furo-cho, Chikusa, Nagoya-shi, Aichi
[4] Hitachi High-Tech Corporation, 882, Ichige, Hitachinaka-shi, Ibaraki
关键词
High-brightness NEA photocathode; Negative electron affinity (NEA); Pulsed electron gun; Scanning electron microscope (SEM); Time-resolved measurement;
D O I
10.1016/j.ultramic.2021.113386
中图分类号
学科分类号
摘要
Temporal changes in carrier relaxations, magnetic switching, and biological structures are known to be in the order of ns. These phenomena can be typically measured by means of an optical-pump & electron-probe method using an electron microscope combined with a pulsed electron source. A photoemission-type pulsed electron gun makes it possible to obtain a short-pulsed electron beam required for high temporal resolution. On the other hand, spatial resolution is restricted by the brightness of the pulsed electron gun used in electron microscopes when a low brightness electron source is used and an irradiation current larger than a certain value is required. Thus, we constructed a prototype pulsed electron gun using a negative electron affinity (NEA) photocathode for time-resolved measurement using a scanning electron microscope (SEM) with high spatiotemporal resolution. In this study, a high-speed detector containing an avalanche photodiode (APD) was used to directly measure waveforms of the pulsed electron beam excited by a rectangular-shape pulsed light with a variable pulse duration in the range of several ns to several μs. The measured waveforms were the same rectangular shape as incident pulsed excitation light. The maximum peak brightness of the pulsed electron beam was 4.2×107 A/m2/sr/V with a pulse duration of 3 ns. This value was larger than that of the continuous electron beam (1.6 × 107 A/m2/sr/V). Furthermore, an SEM image with image sharpness of 6.2 nm was obtained using an SEM equipped with a prototype pulsed electron gun at an acceleration voltage of 3 kV. © 2021
引用
收藏
相关论文
共 50 条
  • [1] Resolution improvement of low-voltage scanning electron microscope by bright and monochromatic electron gun using negative electron affinity photocathode
    Morishita, Hideo
    Ohshima, Takashi
    Kuwahara, Makoto
    Ose, Yoichi
    Agemura, Toshihide
    JOURNAL OF APPLIED PHYSICS, 2020, 127 (16)
  • [2] PULSED NEGATIVE ELECTRON-AFFINITY CATHODE FOR TIME-RESOLVED ELECTRON-MICROSCOPY AND ELECTRON-SPECTROSCOPY
    MACDONALD, NC
    SANFORD, CA
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1988, 135 (08) : C370 - C370
  • [3] A compact electron gun for time-resolved electron diffraction
    Robinson, Matthew S.
    Lane, Paul D.
    Wann, Derek A.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2015, 86 (01):
  • [4] Characterization of a time-resolved electron microscope with a Schottky field emission gun
    Olshin, Pavel K.
    Drabbels, Marcel
    Lorenz, Ulrich J.
    STRUCTURAL DYNAMICS-US, 2020, 7 (05):
  • [5] The evaluation system of Negative Electron Affinity photocathode
    Fu, RG
    Chang, BK
    Qian, YS
    Wang, GH
    Zong, ZY
    APOC 2001: ASIA-PACIFIC OPTICAL AND WIRELESS COMMUNICATIONS: OPTOELECTRONICS, MATERIALS, AND DEVICES FOR COMMUNICATIONS, 2001, 4580 : 614 - 622
  • [6] The design of time-resolved electron microscope
    Li, SH
    Niu, H
    22ND INTERNATIONAL CONGRESS ON HIGH-SPEED PHOTOGRAPHY AND PHOTONICS, 1997, 2869 : 222 - 228
  • [7] Ultrafast miniaturized pulsed electron gun for time-resolved surface measurements
    Epp, Dennis
    Storeck, Gero
    Vogelgesang, Simon
    Sivis, Murat
    Schaefer, Sascha
    Ropers, Claus
    2018 7TH ELECTRONIC SYSTEM-INTEGRATION TECHNOLOGY CONFERENCE (ESTC), 2018,
  • [8] Brightness Measurement of 500kV Electron Microscope Gun
    Shimoyama, Hiroshi
    Ohshita, Akinori
    Maruse, Susumu
    Minamikawa, Yoshihisa
    1600, Oxford University Press (21):
  • [9] IMPROVEMENT IN SCANNING ELECTRON-MICROSCOPE GUN BRIGHTNESS AT LOW KV USING AN INTERMEDIATE EXTRACTION ELECTRODE
    YAMAZAKI, S
    KAWAMOTO, H
    SABURI, K
    NAKATSUKA, H
    BUCHANAN, R
    SCANNING ELECTRON MICROSCOPY, 1984, : 23 - 28
  • [10] Electron beam brightness from negative-electron-affinity photocathodes for scanning electron microscopy application
    Ohshima, T
    Kudo, M
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 2004, 43 (12): : 8335 - 8340