Medical device report analyses from MAUDE: Device and patient outcomes, adverse events, and sex-based differential effects

被引:3
|
作者
Liao, Tsung-Jen [1 ]
Crosby, Lynn [2 ,5 ]
Cross, Kevin [3 ]
Chen, Minjun [1 ]
Elespuru, Rosalie [2 ,4 ]
机构
[1] US FDA, Off Res, Natl Ctr Toxicol Res, Jefferson, AR 72079 USA
[2] US FDA, Off Sci & Engn Labs, Ctr Devices & Radiol Hlth, 10903 New Hampshire Ave, Silver Spring, MD 20993 USA
[3] Instem Inc, Columbus, OH 43217 USA
[4] 3741 Thomas Point Rd, Annapolis, MD 21403 USA
[5] US FDA, Ctr Food Safety & Appl Nutr, Off Analyt & Outreach, College Pk, MD 20740 USA
关键词
MAUDE; Medical devices; Post-market assessment; Patient outcomes; Women's health; CARDIAC PERFORATION;
D O I
10.1016/j.yrtph.2024.105591
中图分类号
DF [法律]; D9 [法律]; R [医药、卫生];
学科分类号
0301 ; 10 ;
摘要
Post-market medical device-associated failures and patient problems are reported in Medical Device Reports (MDRs) to the US Food and Drug Administration. Reports are accessible through Manufacturer and User Facility Device Experience (MAUDE), a database including both required and voluntary submissions. We present an overview of >10 million MDRs received from 2011 to 2021. Approximately 92% of reporting issues represent medical device physical or functional failures, categorized from 1704 codes related to medical device integrity or function. -8% were coded adverse events (AEs). Patient outcomes are reported via 998 patient codes in 19 medical specialties (cardiovascular, orthopedic, etc.). -40% of patient reports indicated "no health consequences"; however, a small number of devices had consistently high AE reports. While overall reports did not exhibit a sex-based dichotomy, -9% of the reported AEs occurred more frequently in females, many of which were related to immune effects. The analyses are subject to uncertainties and potential bias based on data available and data selected for analysis. However, such an overview of post-market MDR data, not previously published, fills a gap in understanding medical device issues and patient-based outcomes related to medical device use. Trends identified may be subjects of additional hypotheses, analysis, and research.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] Adverse Events and Modes of Failure Related to Impella RP: Insights from the Manufacturer and User Facility Device Experience (MAUDE) Database
    Khalid, Nauman
    Rogers, Toby
    Shlofmitz, Evan
    Chen, Yuefeng
    Musallam, Anees
    Khan, Jaffar M.
    Iantorno, Micaela
    Gajanana, Deepakraj
    Hashim, Hayder
    Torguson, Rebecca
    Bernardo, Nelson
    Waksman, Ron
    CARDIOVASCULAR REVASCULARIZATION MEDICINE, 2019, 20 (06) : 503 - 506
  • [22] Sex-based Differences In Outcomes Of Patients Admitted With Heart Failure And Undergoing Left Ventricular Assist Device Implantation
    Udongwo, Ndausung
    Akhlaq, Hira
    Dandu, Sowmya
    Mararenko, Anton
    Odak, Mihir
    James, Dreamy
    Almendral, Jesus
    Heaton, Joseph
    JOURNAL OF CARDIAC FAILURE, 2024, 30 (01) : 250 - 250
  • [23] Patient-reported outcomes and medical device evaluation: from conception to implementation
    Christina M. Webber
    Brittany Caldwell
    Fraser D. Bocell
    Anindita Saha
    Michelle E. Tarver
    Journal of Patient-Reported Outcomes, 6
  • [24] Patient-reported outcomes and medical device evaluation: from conception to implementation
    Webber, Christina M.
    Caldwell, Brittany
    Bocell, Fraser D.
    Saha, Anindita
    Tarver, Michelle E.
    JOURNAL OF PATIENT-REPORTED OUTCOMES, 2022, 6 (01)
  • [25] DEVICE-RELATED AND PATIENT ADVERSE EVENTS ASSOCIATED WITH RADIOFREQUENCY ABLATION AND CRYOABLATION THERAPIES FOR BARRETT'S ESOPHAGUS: A TEN-YEAR ANALYSIS FROM THE MAUDE DATABASE
    Minteer, William
    Chang, Patrick
    Goodman, Morgan
    Gopalkrishnan, Kalpana
    Nguyen, Denis
    Sahakian, Ara
    Buxbaum, James
    Phan, Jennifer
    GASTROINTESTINAL ENDOSCOPY, 2024, 99 (06) : AB406 - AB406
  • [26] Adverse Events After Left Atrial Appendage Closure: Lessons Learned From the Manufacturer and User Facility Device Experience (MAUDE) Database
    Singh, Swaiman
    Okoh, Alexis
    Sossou, Christoph
    Heaton, Joseph
    Sandhu, Mayal
    Chakrabarti, Ritu
    Rao, Rameez
    Tayal, Rajiv
    Wasty, Najam
    Waxman, Sergio
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2019, 74 (13) : B113 - B113
  • [27] Adverse Events After Left Atrial Appendage Closure: Lessons Learned From the Manufacturer and User Facility Device Experience (MAUDE) Database
    Heaton, Joseph
    Okoh, Alexis K.
    Sossou, Christoph
    Singh, Swaiman
    Sandhu, Mayal
    Chakrabarti, Ritu
    Rao, Rameez
    Waxman, Sergio
    Tayal, Rajiv
    Wasty, Najam
    JOURNAL OF INVASIVE CARDIOLOGY, 2020, 32 (08): : E216 - E218
  • [28] ADVERSE EVENTS OF THIRD GENERATION LEFT VENTRICULAR ASSIST DEVICES: INSIGHTS FROM THE MANUFACTURER AND USER FACILITY DEVICE EXPERIENCE (MAUDE) DATABASE
    Heaton, Joseph
    Li, Matthew
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2021, 77 (18) : 645 - 645
  • [29] Adverse Events and Complications Associated With Intrathecal Drug Delivery Systems: Insights From the Manufacturer and User Facility Device Experience (MAUDE) Database
    Goel, Vasudha
    Yang, Yan
    Kanwar, Siddak
    Banik, Ratan K.
    Patwardhan, Amol M.
    Ibrahim, Mohab
    Sivanesan, Eellan
    Shankar, Hariharan
    NEUROMODULATION, 2021, 24 (07): : 1181 - 1189
  • [30] Device- Related Adverse Events and Modes of Failure Related to Renal Denervation Systems: Insights From Analysis of the FDA MAUDE Database
    Jasti, Manasa
    Nanjundappa, Aravinda
    Paul, Timir
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2024, 84 (18) : B159 - B159