PEAK DISTORTION EFFECTS OF SURAMIN DUE TO LARGE SYSTEM PEAKS IN BIOANALYSIS USING ION-PAIR ADSORPTION CHROMATOGRAPHY

被引:16
|
作者
FORNSTEDT, T
机构
[1] Department of Analytical Pharmaceutical Chemistry, Uppsala University Biomedical Centre, S-751 23 Uppsala
来源
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS | 1993年 / 612卷 / 01期
关键词
D O I
10.1016/0378-4347(93)80376-F
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Sample injections in ion-pair adsorption chromatography lead to changes in the established equilibria of the eluent components at the top of the column, which in turn lead to migrating concentration changes in the column (system peaks). Large system peaks contain big concentration deviations of the mobile phase components compared with the bulk composition in the eluent. Analyte peaks are distorted upon combined elution with large system peaks. In bioanalysis, samples that deviate considerably from the eluent are often injected, resulting in large system peaks. An illustrative example of such analyte peak distortions in bioanalysis is given for the case of suramin, and guidelines are given for avoiding the effects.
引用
收藏
页码:137 / 144
页数:8
相关论文
共 23 条