Effect of the bone heterogeneity on the dose prescription in orthovoltage radiotherapy: A Monte Carlo study

被引:7
|
作者
Chow, James C. L. [1 ,2 ,3 ]
Grigorov, Grigor N. [4 ]
机构
[1] Univ Toronto, Princess Margaret Hosp, Radiat Med Program, Toronto, ON M5G 2M9, Canada
[2] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5G 2M9, Canada
[3] Ryerson Univ, Dept Phys, Toronto, ON M5B 2K3, Canada
[4] Grand River Reg Canc Ctr, Dept Med Phys, Kitchener, ON N2G IG3, Canada
关键词
Orthovoltage photon beam; Orthovoltage radiation therapy; Monte Carlo simulation; Bone backscatter and surface dose;
D O I
10.1016/j.rpor.2011.09.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: In orthovoltage radiotherapy, since the dose prescription at the patient's surface is based on the absolute dose calibration using water phantom, deviation of delivered dose is found as the heterogeneity such as bone present under the patient's surface. Aim: This study investigated the dosimetric impact due to the bone heterogeneity on the surface dose in orthovoltage radiotherapy. Materials and methods: A 220 kVp photon beam with field size of 5 cm diameter, produced by a Gulmay D3225 orthovoltage X-ray machine was modeled by the BEAMnrc. Phantom containing water (thickness = 1-5 mm) on top of a bone (thickness = 1 cm) was irradiated by the 220 kVp photon beam. Percentage depth dose (PDD), surface dose and photon energy spectrum were determined using Monte Carlo simulations (the BEAMnrc code). Results: PDD results showed that the maximum bone dose was about 210% higher than the surface dose in the phantoms with different thicknesses of water. Surface dose was found to be increased in the range of 2.5-3.7%, when the distance between the phantom surface and bone was increased in the range of 1-5 mm. The increase of surface dose was found not to follow the increase of water thickness, and the maximum increase of surface dose was found at the thickness of water equal to 3 mm. Conclusions: For the accepted total orthovoltage radiation treatment uncertainty of 5%, a neglected consideration of the bone heterogeneity during the dose prescription in the sites of forehead, chest wall and kneecap with soft tissue thickness = 1-5 mm would cause more than two times of the bone dose, and contribute an uncertainty of about 2.5-3.7% to the total uncertainty in the dose delivery. (C) 2011 Greater Poland Cancer Centre, Poland. Published by Elsevier Urban & Partner Sp. z.o.o. All rights reserved.
引用
收藏
页码:38 / 43
页数:6
相关论文
共 50 条
  • [1] Evaluation of Dosimetric Effect of Bone Scatter on Nanoparticle-Enhanced Orthovoltage Radiotherapy: A Monte Carlo Phantom Study
    Sadiq, Afia
    Chow, James C. L.
    [J]. NANOMATERIALS, 2022, 12 (17)
  • [2] Depth Dose Enhancement in Orthovoltage Nanoparticle-Enhanced Radiotherapy: A Monte Carlo Phantom Study
    Chow, James C. L.
    Jubran, Sama
    [J]. MICROMACHINES, 2023, 14 (06)
  • [3] Dosimetric dependence on bone backscatter in orthovoltage radiotherapy: A Monte Carlo photon fluence spectral study
    Chow, J.
    Grigor, G.
    [J]. MEDICAL PHYSICS, 2014, 41 (08) : 17 - 17
  • [4] A Monte Carlo study on dose distribution of an orthovoltage radiation therapy system
    Iseri, Toshie
    Tanabe, Yoshinori
    Onizuka, Ryouta
    Torigoe, Yuri
    Horikirizono, Hiro
    Itamoto, Kazuhito
    Sunahara, Hiroshi
    Itoh, Harumichi
    Tani, Kenji
    Nakaichi, Munekazu
    [J]. PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (02) : 623 - 632
  • [5] A Monte Carlo study on dose distribution of an orthovoltage radiation therapy system
    Toshie Iseri
    Yoshinori Tanabe
    Ryouta Onizuka
    Yuri Torigoe
    Hiro Horikirizono
    Kazuhito Itamoto
    Hiroshi Sunahara
    Harumichi Itoh
    Kenji Tani
    Munekazu Nakaichi
    [J]. Physical and Engineering Sciences in Medicine, 2023, 46 : 623 - 632
  • [6] Effect of dental restorations and prostheses on radiotherapy dose distribution: a Monte Carlo study
    Chin, David W. H.
    Treister, Nathaniel
    Friedland, Bernard
    Cormack, Robert A.
    Tishler, Roy B.
    Makrigiorgos, G. Mike
    Court, Laurence E.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (01): : 80 - 89
  • [7] STEREOTACTIC RADIOTHERAPY FOR LUNG TUMOURS: TREATMENT PRESCRIPTION USING MONTE CARLO DOSE CALCULATION
    Bolt, R.
    Van der Schaaf, A.
    Hollander, M.
    Widder, J.
    van't Veld, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S116 - S116
  • [8] Optimizing dose prescription in stereotactic body radiotherapy for lung tumours using Monte Carlo dose calculation
    Widder, Joachim
    Hollander, Miranda
    Ubbels, Jan F.
    Bolt, Rene A.
    Langendijk, Johannes A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 94 (01) : 42 - 46
  • [9] Dose distribution in bone-Monte Carlo study
    Steiner, M.
    Viziblova, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S223 - S223
  • [10] Study of Dose Calculation Using Monte Carlo Method in Radiotherapy
    Gao, C.
    Chi, Z.
    Liu, D.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)