SPOILING OF TRANSVERSE MAGNETIZATION IN STEADY-STATE SEQUENCES

被引:225
|
作者
ZUR, Y
WOOD, ML
NEURINGER, LJ
机构
[1] MIT,FRANCIS BITTER NATL MAGNET LAB,CAMBRIDGE,MA 02139
[2] NEW ENGLAND MED CTR HOSP,DEPT RADIAT ONCOL,BOSTON,MA 02111
[3] TUFTS UNIV,BOSTON,MA 02111
关键词
D O I
10.1002/mrm.1910210210
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A detailed analysis is presented of a method to eliminate transverse magnetization prior to each rf excitation in pulse sequences with TR < T2. It is shown that artifact‐free images with high T1 contrast can be obtained only if a phase shift that is incremented during each TR interval is applied to the transverse magnetization. Computer simulations are used to show that when this phase increment is 117°, the steady‐state transverse magnetization prior to each rf pulse is nulled over a wide range of T1, T2, and rf tip angles, resulting in optimal T1, contrast. Such nulling of steady‐state transverse magnetization cannot be obtained by using large gradient pulses, or gradients of random or linearly incremented amplitude. Images of phantoms and human subjects confirm the theoretical predictions. © 1991 Academic Press, Inc. Copyright © 1991 Wiley‐Liss, Inc., A Wiley Company