STIMULATION CHARACTERISTICS, COMPLICATIONS, AND EFFICACY OF SPINAL-CORD STIMULATION SYSTEMS IN PATIENTS WITH REFRACTORY ANGINA - A PROSPECTIVE FEASIBILITY STUDY

被引:0
|
作者
DEJONGSTE, MJL
NAGELKERKE, D
HOOYSCHUUR, CM
JOURNEE, HL
MEYLER, PWJ
STAAL, MJ
DEJONGE, P
LIE, KI
机构
[1] MEDTRONIC INC, EINDHOVEN, NETHERLANDS
[2] UNIV GRONINGEN HOSP, DEPT NEUROSURG, 9700 RB GRONINGEN, NETHERLANDS
[3] UNIV GRONINGEN HOSP, DEPT ANESTHESIOL, 9700 RB GRONINGEN, NETHERLANDS
来源
关键词
ANGINA; FOLLOW-UP SPINAL CORD STIMULATION CHARACTERISTICS;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: In a prospective study with a 1-year follow-up we evaluated: (1) the feasibility of a method for the adjustment of spinal cord stimulator (SCS) parameters, (2) complications of SCS, and (3) efficacy of SCS. Methods: In patients receiving an SCS for severe angina unresponsive to standard therapies, SCS characteristics were evaluated within 1 week and at 4, 14, 26, and 52 weeks after SCS treatment. Step-by-step adjustment of pulse output parameters was performed at the electrode configuration at which paresthesias occurred (''sensory threshold''), covered the anginal area (''adjusted setting''), or provoked pain (''motor threshold''). In addition, the number of anginal attacks and intake of glyceryl trinitrate (GTN) tablets were recorded at regular intervals. Results: Twenty-two patients with either a bipolar (14) or a unipolar (8) system were evaluated. In the 14 patients with bipolar systems, alteration in paresthesias required 26 reprogrammings of the configuration. in the eight patients with bipolar systems who completed the followup without lead dislocation, the mean ''sensory threshold'' was 3.3 V (1.7-5.6), the mean ''adjusted stimulation'' output was 4.5 V (2.8-7.6), and the mean ''motor threshold'' was 4.9 V (2.8-7.7) after 4 weeks SCS. The mean stimulation duration per 24 hours was 14.2% (5%-24%), and the mean standardized impedance was 821 Omega (748-893) after 4 weeks SCS. The unipolar group demonstrated comparable results. After 1-year follow-up the parameters had not changed significantly. During the I-year followup, 6 of 22 patients experienced lead dislocation that required surgery. In all patients, anginal attacks (P < 0.003) and GTN intake (P < 0.005) were reduced significantly with SCS. The effect lasted during the 1 year. Conclusions: During a 1-year follow-up, the stimulation parameters did not change significantly in the 16 patients without lead dislocations. Our standardized method appears to be feasible for followup of SCS. Moreover, SCS seems to be an effective adjuvant therapy for intractable angina, despite a relatively frequent dislocation of the electrode.
引用
收藏
页码:1751 / 1760
页数:10
相关论文
共 50 条
  • [1] Spinal cord Stimulation in Refractory Angina
    Rojas Borda, Andres Felipe
    Aldana Bitar, Jairo Alejandro
    Acevedo Gonzalez, Juan Carlos
    UNIVERSITAS MEDICA, 2014, 55 (03): : 311 - 327
  • [2] Spinal-cord stimulation in management of angina
    Brodison, A
    Chauhan, A
    LANCET, 1999, 354 (9192): : 1748 - 1749
  • [3] Longevity and costs of spinal cord stimulation systems in patients with refractory angina pectoris
    Jessurun, GAJ
    TenVaarwerk, IAM
    Tulner, SAF
    Schonhuth, CCP
    Staal, MJ
    DeJongste, MJL
    PACEMAKER LEADS 1997, 1997, : 613 - 619
  • [4] Spinal Cord Stimulation for Refractory Angina Pectoris
    Lee, Seong Heon
    Jeong, Hye Jin
    Jeong, Sin Ho
    Lee, Hyung Gon
    Choi, Jeong Il
    Yoon, Myung Ha
    Kim, Woong Mo
    KOREAN JOURNAL OF PAIN, 2012, 25 (02): : 121 - 125
  • [5] Spinal Cord Stimulation for Refractory Angina Pectoris
    Shabalov, Vladimir
    Murtazin, Vladimir
    Ashurkov, Andrei
    Duishobaev, Abdirakman
    Krivoshapkin, Alexei
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2013, 91 : 272 - 272
  • [6] Spinal Cord Stimulation for Refractory Angina Pectoris
    Hale, Jason
    Bailey-Classen, Ashley
    Cheng, Jianguo
    PAIN MEDICINE, 2020, 21 (01) : 198 - 200
  • [7] Feasibility of Spinal Cord Stimulation in Patients With Refractory Angina Pectoris and a Cardiac Implanted Electronic Device
    Vervaat, Fabienne E.
    van Suijlekom, Hans
    Wijnbergen, Inge
    NEUROMODULATION, 2023, 26 (08): : 1867 - 1875
  • [8] PRELIMINARY-RESULTS OF A RANDOMIZED STUDY ON THE CLINICAL EFFICACY OF SPINAL-CORD STIMULATION FOR REFRACTORY SEVERE ANGINA-PECTORIS
    DEJONGSTE, MJL
    STAAL, MJ
    ACTA NEUROCHIRURGICA, 1993, : 161 - 164
  • [9] EFFICACY OF SPINAL-CORD STIMULATION AS ADJUVANT THERAPY FOR INTRACTABLE ANGINA-PECTORIS - A PROSPECTIVE, RANDOMIZED CLINICAL-STUDY
    DEJONGSTE, MJL
    HAUTVAST, RWM
    HILLEGE, HL
    LIE, KI
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (07) : 1592 - 1597
  • [10] The efficacy of spinal cord stimulation in angina pectoris
    Börjesson, M
    Mannheimer, C
    MANAGEMENT OF ACUTE AND CHRONIC PAIN: THE USE OF THE TOOLS OF THE TRADE, 2000, : 683 - 689