BIOMECHANICAL STUDY OF C1-C2 POSTERIOR ARTHRODESIS TECHNIQUES

被引:62
|
作者
HAJEK, PD [1 ]
LIPKA, J [1 ]
HARTLINE, P [1 ]
SAHA, S [1 ]
ALBRIGHT, JA [1 ]
机构
[1] LOUISIANA STATE UNIV,MED CTR,DEPT ORTHOPAED SURG,SHREVEPORT,LA 71105
关键词
BROOKS FUSION; HALIFAX INTERLAMINAR CLAMPS; GALLIE FUSION;
D O I
10.1097/00007632-199302000-00001
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Fifteen cervical spines from cadavers were used to compare the rotational and translational stability of the Brooks fusion, a fusion construct using Halifax interlaminar clamps, and the Gallie fusion. The Brooks and Halifax clamp constructs exhibited significantly greater rotational and translational stiffness than the Gallie construct (P < 0.001). The Halifax clamp construct exhibited greater rotational stiffness and equal translational stiffness when compared with the Brooks construct (P < 0.05). The Brooks and Halifax fixation constructs provided superior fixation but presented technical challenges. The Gallie construct is less technically demanding but provides less stable fixation.
引用
收藏
页码:173 / 177
页数:5
相关论文
共 50 条
  • [42] Odontoid plate fixation without C1-C2 arthrodesis: Biomechanical testing of a novel surgical technique and comparison to the conventional screw fixation procedure
    Platzer, Patrick
    Eipeldauer, Stefan
    Vecsei, Vilmos
    CLINICAL BIOMECHANICS, 2010, 25 (07) : 623 - 627
  • [43] C1-C2 Posterior Fixation: Are 4 Screws Better Than 2? COMMENTS
    Sonntag, Volker K. H.
    Harrop, James S.
    NEUROSURGERY, 2012, 71 : 95 - 95
  • [44] Posterior C1-C2 transarticular screw fixation for instability of the C0-C2 complex
    Martín-Ferrer, S
    Rimbau, J
    Joly, MC
    Teruel, J
    Pont, J
    NEUROCIRUGIA, 2000, 11 (01): : 43 - 49
  • [45] Inadvertent C2-C3 union after C1-C2 posterior fusion in adults
    Takeuchi, K
    Yokoyama, T
    Aburakawa, S
    Ueyama, K
    Ito, J
    Sannohe, A
    Okada, A
    Toh, S
    EUROPEAN SPINE JOURNAL, 2006, 15 (03) : 270 - 277
  • [46] Atlantoaxial arthrodesis using C1-C2 transarticular screw fixation in a case of Morquio syndrome
    Arvind G. Kulkarni
    Siddharth M. Shah
    Indian Journal of Orthopaedics, 2011, 45 : 470 - 472
  • [47] A Computed Tomographic Study of Vietnamese C1-C2 Morphology for Atlantoaxial Crew Fixation Techniques
    Nguyen Duy Hung
    Nguyen Minh Duc
    Le Viet Dung
    Than Van Sy
    Le Thanh Dung
    Nguyen Duy Hue
    JOURNAL OF CLINICAL IMAGING SCIENCE, 2020, 10
  • [48] Atlantoaxial arthrodesis using C1-C2 transarticular screw fixation in a case of Morquio syndrome
    Kulkarni, Arvind G.
    Shah, Siddharth M.
    INDIAN JOURNAL OF ORTHOPAEDICS, 2011, 45 (05) : 470 - 472
  • [49] Intraoperative Doppler ultrasound as a means of preventing vertebral artery injury during Goel and Harms C1-C2 posterior arthrodesis: technical note
    Lofrese, Giorgio
    Cultrera, Francesco
    Visani, Jacopo
    Nicassio, Nicola
    Ibn Essayed, Walid
    Donati, Roberto
    Cavallo, Michele Alessandro
    De Bonis, Pasquale
    JOURNAL OF NEUROSURGERY-SPINE, 2019, 31 (06) : 824 - 830
  • [50] Treatment of iatrogenic vertebral artery injury during C1-C2 arthrodesis with a covered stent
    Bram, Richard
    Mehta, Ankit, I
    Atwal, Gursant
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2024,