COVARIANCE ANALYSIS AND FITTING OF GERMANIUM GAMMA-RAY DETECTOR EFFICIENCY CALIBRATION DATA

被引:60
|
作者
GERALDO, LP [1 ]
SMITH, DL [1 ]
机构
[1] ARGONNE NATL LAB,DIV ENGN PHYS,ARGONNE,IL 60439
来源
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT | 1990年 / 290卷 / 2-3期
关键词
D O I
10.1016/0168-9002(90)90569-R
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The measurement of neutron-induced reaction cross sections often requires detection of either prompt gamma rays or gamma rays emitted during the decay of radio nuclides. This paper describes the analytical procedures we use to generate calibrations for germanium gamma-ray detector full-energy-peak efficiency ε{lunate} versus photon energy E, and to predict the associated uncertainties. Our method, which involves fitting a parameterized regression formula to measured data by the principle of least squares, is widely applicable beyond the specific problem considered here. It differs from most commonly used methods in that comprehensive use is made of available information on all known sources of random and systematic (correlated) error associated with the calibration process. This is accomplished through the formation and application of a calibration-data covariance matrix. Objective prediction of the errors in subsequently derived quantities (e.g., detector efficiencies at energies not directly represented in the calibration data) is then achieved through error propagation. Specifically, this paper discusses our experiences in fitting the particular empirical formula In ε{lunate} = Σk-1m pk(InE)k-1 to measured calibration data at gamma-ray energies from just above 200 keV to several MeV. A numerical example is provided to demonstrate the utility of this approach. © 1990.
引用
收藏
页码:499 / 508
页数:10
相关论文
共 50 条
  • [21] A germanium orthogonal strip detector system for gamma-ray imaging
    Hull, EL
    Burks, MT
    Cork, CP
    Craig, W
    Eckels, D
    Fabris, L
    Lavietes, A
    Luke, PN
    Madden, NW
    Pehl, RH
    Ziock, KP
    HARD X-RAY AND GAMMA-RAY DETECTOR PHYSICS III, 2001, 4507 : 132 - 140
  • [22] RADIOACTIVE STANDARDS FOR CALIBRATION OF GERMANIUM DETECTORS FOR GAMMA-RAY SPECTROMETRY
    DEBERTIN, K
    SCHOTZIG, U
    WALZ, KF
    WEISS, HM
    PTB-MITTEILUNGEN, 1977, 87 (01): : 22 - 26
  • [23] THE ROLE OF COVARIANCE IN THE ANALYSIS OF GAMMA-RAY SPECTRA
    GRAY, PW
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1986, 102 (01): : 59 - 67
  • [24] A nuclear medicine gamma-ray detector based on germanium strip detector technology
    Hall, CJ
    Helsby, WI
    Lewis, RA
    Nolan, P
    Boston, A
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 513 (1-2): : 47 - 50
  • [25] Covariance analysis of efficiency calibration of NaI(Tl) detector
    Ryngnga, Jodie T.
    Jyrwa, Betylda
    MATERIALS TODAY-PROCEEDINGS, 2022, 65 : 2778 - 2780
  • [26] Contribution of a germanium detector in mobile gamma-ray spectrometry. Spectral analysis and performance
    Gutierrez, S
    Guillot, L
    Bourgeois, C
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 482 (1-2): : 425 - 433
  • [27] Contribution of a germanium detector in mobile gamma-ray spectrometry. Spectral analysis and performance
    Gutierrez, S.
    Guillot, L.
    Bourgeois, C.
    2002, Elsevier (481) : 1 - 3
  • [28] A GAMMA-RAY DETECTOR WITH LOW EFFICIENCY FOR FAST NEUTRONS
    BISHOP, GR
    PRESTON, G
    WESTHEAD, JM
    HALBAN, H
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1953, 24 (05): : 395 - 396
  • [29] EFFICIENCY CALIBRATION AND SUMMATION EFFECTS IN GAMMA-RAY SPECTROMETRY
    ERTEN, HN
    AKSOYOGLU, S
    GOKTURK, H
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1988, 125 (01): : 3 - 10
  • [30] On the preparation of efficiency calibration standards for gamma-ray spectrometers
    Taskaeva, M
    Taskaev, E
    Penev, I