Tracheal Rupture due to Diffusion of Nitrous Oxide to Cuff of High-Volume, Low-Pressure Intubation Tube

被引:0
|
作者
Atalay, Canan [1 ]
Aykan, Seyda [1 ]
Can, Abdullah [1 ]
Dogan, Nazim [1 ]
机构
[1] Ataturk Univ, Fac Med, Dept Anesthesiol & Reanimat, TR-25070 Erzurum, Turkey
来源
EURASIAN JOURNAL OF MEDICINE | 2009年 / 41卷 / 02期
关键词
Tracheal rupture; High-volume; low-pressure endotracheal tube; Nitrous oxide;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Tracheal rupture is a rare complication of endotracheal intubation. Risk factors include short neck, repeated attempts due to failed intubation, inappropriate stylus, over-inflation of the cuff, poor positioning of the tube, inappropriate tube size, weakened membrane structure due to steroid use, chronic obstructive pulmonary disease, tracheomalacia, kyphosis, and use of nitric oxide during the operation. In this article, we suggest that high-volume, low-pressure tubes may not always provide a low-pressure effect and could rupture due to reduced tracheal perfusion pressure and ischemic damage upon over-inflation.
引用
收藏
页码:136 / 139
页数:4
相关论文
共 50 条
  • [1] The tracheal tube with a high-volume, low-pressure cuff at various airway inflation pressures
    Inada, T
    Uesugi, F
    Kawachi, S
    Inada, K
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 1998, 15 (06) : 629 - 632
  • [2] DILATATION OF TRACHEA IN NASO-TRACHEAL LONG-TERM INTUBATION WITH HIGH-VOLUME LOW-PRESSURE CUFF - TECHNICAL COMPLICATION
    SKARVAN, K
    DITTMANN, M
    LEHMANN, K
    WOLFF, G
    SCHWEIZERISCHE MEDIZINISCHE WOCHENSCHRIFT, 1977, 107 (45) : 1635 - 1635
  • [3] Improving the shape and compliance characteristics of a high-volume, low-pressure cuff improves tracheal seal
    Young, PJ
    Blunt, MC
    BRITISH JOURNAL OF ANAESTHESIA, 1999, 83 (06) : 887 - 889
  • [4] ASPIRATION AROUND HIGH-VOLUME, LOW-PRESSURE ENDOTRACHEAL CUFF
    SANDERSON, P
    BRITISH MEDICAL JOURNAL, 1981, 283 (6305): : 1546 - 1547
  • [5] ASPIRATION AROUND HIGH-VOLUME, LOW-PRESSURE ENDOTRACHEAL CUFF
    MEHTA, S
    BRITISH MEDICAL JOURNAL, 1982, 284 (6309): : 115 - 116
  • [6] ASPIRATION AROUND HIGH-VOLUME, LOW-PRESSURE ENDOTRACHEAL CUFF
    MACRAE, W
    WALLACE, P
    BRITISH MEDICAL JOURNAL, 1981, 283 (6301): : 1220 - 1220
  • [7] A low-volume, low-pressure tracheal tube cuff reduces pulmonary aspiration
    Young, PJ
    Pakeerathan, S
    Blunt, MC
    Subramanya, S
    CRITICAL CARE MEDICINE, 2006, 34 (03) : 632 - 639
  • [8] EFFECT OF TRACHEAL DILATATION AND RUPTURE ON MECHANICAL VENTILATION USING A LOW-PRESSURE CUFF TUBE
    LUNA, CM
    LEGARRETA, G
    ESTEVA, H
    LAFFAIRE, E
    JOLLY, EC
    CHEST, 1993, 104 (02) : 639 - 640
  • [9] An in-vitro study to evaluate high-volume low-pressure endotracheal tube cuff deflation dynamics
    Marti, Joan D.
    Li Bassi, Gianluigi
    Isetta, Valentina
    Lazaro, Miguel R.
    Aguilera-Xiol, Eli
    Comaru, Talitha
    Battaglini, Denise
    Meli, Andrea
    Ferrer, Miguel
    Navajas, Daniel
    Pelosi, Paolo
    Chiumello, Davide
    Torres, Antoni
    Farre, Ramon
    MINERVA ANESTESIOLOGICA, 2019, 85 (08) : 846 - 853
  • [10] HIGH-VOLUME, LOW-PRESSURE CUFFS - ARE THEY ALWAYS LOW-PRESSURE
    GUYTON, D
    BANNER, MJ
    KIRBY, RR
    CHEST, 1991, 100 (04) : 1076 - 1081