PHENYLKETONURIA SCREENING WITH A FLUOROMETRIC MICROPLATE ASSAY

被引:0
|
作者
LUBENOW, N [1 ]
DIEPENBROCK, F [1 ]
SCHICKLING, H [1 ]
BOCK, D [1 ]
HECKLER, R [1 ]
SANDER, J [1 ]
机构
[1] STAATLICHES MED UNTERSUCHUNGSAMT HANNOVER,ROESEBECKSTR 4-6,D-30449 HANNOVER,GERMANY
关键词
D O I
暂无
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
A fluorometric assay in microtitre plates for the screening of phenylketonuria was evaluated and adapted to a neonatal screening programme. Using this assay, it is possible to determine quantitatively the phenylalanine concentration in dried blood spots on filter paper. The test exhibited a linear calibration curve with a good slope as well as sufficient precision and accuracy in the statistical analysis. Interference by other amino acids and antibiotics was not observed. Only elevated concentrations of leucine interfered to a small degree. The phenylalanine concentration in dried blood spots of 13 phenylketonuria patients correlated to that in serum. 7381 dried blood samples of newborn infants were tested simultaneously by both the fluorometric and the Guthrie test. The results did not show significant differences. We screened 29182 newborns using the fluorometric assay and an online data processing programme. The internal repetition rate was 0.64%, the external recall rate 0.15%. False negative results were not observed. In December 1991 the fluorometric method replaced the Guthrie test in our routine programme for phenylketonuria screening, and was introduced as a follow up test for phenylketonuria patients.
引用
收藏
页码:525 / 528
页数:4
相关论文
共 50 条
  • [41] UTILISATION OF AN AUTOMATED FLUOROMETRIC METHOD FOR DETECTING PHENYLKETONURIA
    COLLOMBEL, C
    DORCHIE, C
    BALTASSAT, P
    COTTE, J
    ANNALES DE BIOLOGIE CLINIQUE, 1969, 27 (10-1) : 671 - +
  • [42] A fluorometric assay for acetylcholinesterase activity and inhibitor screening with carbon quantum dots
    Qian, Zhaosheng
    Chai, Lujing
    Tang, Cong
    Huang, Yuanyuan
    Chen, Jianrong
    Feng, Hui
    SENSORS AND ACTUATORS B-CHEMICAL, 2016, 222 : 879 - 886
  • [43] A fluorometric assay for high-throughput screening targeting nicotinamide phosphoribosyltransferase
    Zhang, Ruo-Yu
    Qin, Ye
    Lv, Xiao-Qun
    Wang, Pei
    Xu, Tian-Ying
    Zhang, Lei
    Miao, Chao-Yu
    ANALYTICAL BIOCHEMISTRY, 2011, 412 (01) : 18 - 25
  • [44] SCREENING FOR PHENYLKETONURIA
    HOLTZMAN, NA
    MCCABE, ERB
    CUNNINGHAM, GC
    BERRY, HK
    NEW ENGLAND JOURNAL OF MEDICINE, 1981, 304 (21): : 1300 - 1300
  • [45] SCREENING FOR PHENYLKETONURIA
    KOMROWER, GM
    LANCET, 1975, 1 (7902): : 328 - 328
  • [46] SCREENING FOR PHENYLKETONURIA
    CRAWFURD, MD
    BRITISH MEDICAL JOURNAL, 1974, 1 (5901): : 241 - 241
  • [47] A new automatized fluorometric assay for anti-Leishmania drug screening
    Le Pape, P
    Pagniez, F
    Abdala, H
    ACTA PARASITOLOGICA, 2002, 47 (01) : 79 - 81
  • [48] SCREENING FOR PHENYLKETONURIA
    CUSWORTH, DC
    BRITISH MEDICAL JOURNAL, 1968, 4 (5626): : 324 - &
  • [49] SCREENING FOR PHENYLKETONURIA
    OBRIEN, JM
    PHILPOTT, MG
    BRITISH MEDICAL JOURNAL, 1968, 4 (5629): : 515 - &
  • [50] PHENYLKETONURIA SCREENING
    SNYDERMAN, SE