POTENTIOMETRIC BIOSENSOR IN DOUBLE MATRIX MEMBRANE TECHNOLOGY

被引:15
|
作者
EGGENSTEIN, C
BORCHARDT, M
DUMSCHAT, C
GRUNDIG, B
CAMMANN, K
SPENER, F
KNOLL, M
机构
[1] Institut für Chemo- und Biosensorik Münster, D 48149 Münster
来源
BIOSENSORS & BIOELECTRONICS | 1995年 / 10卷 / 6-7期
关键词
POTENTIOMETRIC BIOSENSOR; UREA SENSOR; DISPOSABLE BIOSENSOR;
D O I
10.1016/0956-5663(95)96935-R
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A throw-away urea sensor is described. A pH-sensitive disposable sensor constructed using double matrix membrane technology was used as a transducer. Double matrix membrane technology describes a process where an ion-sensitive polymer matrix membrane is formed in the presence of an additional inert matrix for to improve the reproducibility of sensor production. In this case one-side silver-coated filter paper was used as the inert matrix which was encapsulated by a heat sealing film. A pH-sensitive membrane was casted into this filter paper and on the surface of this membrane a urease containing layer was deposited. The sensor responded to between 10(-2) mol/l and 10(-4) mol/l urea, the response slope was 32.7 mV/decade. The concentration range in which the sensor responds is important for clinical analysis. The main advantages of this pH-based urea electrode include miniature size as well as simple and low cost fabrication in a batch process.
引用
收藏
页码:595 / 600
页数:6
相关论文
共 50 条
  • [31] Reed Membrane as a Novel Immobilization Matrix for the Development of an Optical Phenol Biosensor
    Li, Yongjin
    CURRENT ANALYTICAL CHEMISTRY, 2020, 16 (03) : 316 - 320
  • [32] PVC matrix membrane sensor for potentiometric determination of triphenyltetrazolium chloride and ascorbic acid
    Mostafa, Gamal Abdel-Hafiz
    ANNALI DI CHIMICA, 2007, 97 (11-12) : 1247 - 1256
  • [33] A DOUBLE-MEMBRANE ION-SELECTIVE ELECTRODE FOR THE POTENTIOMETRIC DETERMINATION OF POTASSIUM
    DIAZ, C
    VIDAL, JC
    GALBAN, J
    URARTE, ML
    LANAJA, J
    MICROCHEMICAL JOURNAL, 1989, 39 (03) : 289 - 297
  • [34] A novel inhibition based biosensor using urease nanoconjugate entrapped biocomposite membrane for potentiometric glyphosate detection
    Vaghela, Chetana
    Kulkarni, Mohan
    Haram, Santosh
    Aiyer, Rohini
    Karve, Meena
    INTERNATIONAL JOURNAL OF BIOLOGICAL MACROMOLECULES, 2018, 108 : 32 - 40
  • [35] Porous silicon based potentiometric triglyceride biosensor
    Reddy, RRK
    Chadha, A
    Bhattacharya, E
    BIOSENSORS & BIOELECTRONICS, 2001, 16 (4-5): : 313 - 317
  • [36] Potentiometric biosensor for urinary glucose level monitoring
    Kormos, F
    Sziráki, L
    Tarsiche, I
    LABORATORY ROBOTICS AND AUTOMATION, 2000, 12 (06) : 291 - 295
  • [37] Improved performance of the potentiometric biosensor for the determination of creatinine
    Dallerup Rasmussen, Claus
    Andersen, Jens E. T.
    Zachau-Christiansen, Birgit
    ANALYTICAL LETTERS, 2007, 40 (01) : 39 - 52
  • [38] The Potentiometric Biosensor Acetylcholine Esterase as a Model System
    Stephan Reher
    Yvonne Lepka
    Georg Schwedt
    Microchimica Acta, 2002, 140 : 15 - 20
  • [39] Urease immobilized potentiometric biosensor for determination of urea
    Sihombing, Kawan
    Tamba, Monica Chrisdayanti
    Marbun, Winry Sulistiani
    Situmorang, Manihar
    INDIAN JOURNAL OF CHEMISTRY SECTION A-INORGANIC BIO-INORGANIC PHYSICAL THEORETICAL & ANALYTICAL CHEMISTRY, 2018, 57 (02): : 175 - 180
  • [40] Development of potentiometric biosensor for diagnosis of prostate cancer
    Altunkok, Nazire
    Unluer, Ozlem Bicen
    Ozkutuk, Ebru Birlik
    Ersoz, Arzu
    MATERIALS SCIENCE AND ENGINEERING B-ADVANCED FUNCTIONAL SOLID-STATE MATERIALS, 2021, 263