FAILURES OF EPICARDIAL AND TRANSVENOUS LEADS FOR IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS

被引:21
|
作者
MATTKE, S [1 ]
MULLER, D [1 ]
MARKEWITZ, A [1 ]
KAULBACH, H [1 ]
SCHMOCKEL, M [1 ]
DORWARTH, U [1 ]
HOFFMANN, E [1 ]
STEINBECK, G [1 ]
机构
[1] UNIV MUNICH,DEPT CARDIAC SURG,D-81366 MUNICH,GERMANY
关键词
D O I
10.1016/0002-8703(95)90206-6
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Lead failures constitute a major risk for patients with an implantable cardioverter defibrillator (ICD). However, data about the incidence and patterns of ICD-iead failures in a larger population are lacking. We analyzed the short-term and midterm performance of 27 epicardial and 103 nonthoracotomy ICD-lead systems during a follow-up period of 36 +/- 21 months and 22 +/- 10 months, respectively (p < 0.05). The failure rate was 5 (19%) of 27 In the epicardial and 6 (6%) of 103 in the nonthoracotomy group (p < 0.05). The most common symptom was erroneous detection of ventricular fibrillation from artifact sensing in five patients. Two patients had to be resuscitated because of failure to defibrillate. toss of pacing and loss of sensing were seen In two patients. Only two asymptomatic lead fractures could be diagnosed on routine radiograph. In conclusion, there was a considerable rate of lead failures, especially in epicardial systems. Longterm studies addressing the longevity of ICD leads, mechanisms of their failures, and Improved diagnostic facilities are important to further increase the safety of this therapeutic approach.
引用
收藏
页码:1040 / 1044
页数:5
相关论文
共 50 条
  • [21] Implantable cardioverter-defibrillators in children
    Blom, Nico A.
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2008, 31 : S32 - S34
  • [22] Deactivation of implantable cardioverter-defibrillators
    Wan, Darryl
    Chakrabarti, Santabhanu
    [J]. CANADIAN MEDICAL ASSOCIATION JOURNAL, 2021, 193 (23) : E852 - E852
  • [23] The Deactivation of Implantable Cardioverter-Defibrillators
    Carlsson, Jorg
    Paul, Norbert W.
    Dann, Matthias
    Neuzner, Joerg
    Pfeiffer, Dietrich
    [J]. DEUTSCHES ARZTEBLATT INTERNATIONAL, 2012, 109 (33-34): : 535 - U14
  • [24] Driving and implantable cardioverter-defibrillators
    Grubb, Blair P.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 50 (23) : 2241 - 2242
  • [25] Implantable cardioverter-defibrillators - Reply
    McClellan, MB
    Tunis, SR
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2005, 352 (19): : 2025 - 2025
  • [26] Implantable cardioverter-defibrillators and complications
    Sinclair, CT
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2004, 350 (07): : 734 - 734
  • [27] Implantable cardioverter-defibrillators in children
    DeWitt, Elizabeth S.
    Abrams, Dominic J.
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD, 2015, 100 (03) : 265 - 270
  • [28] MORTALITY AND IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS
    MOWER, MM
    [J]. CIRCULATION, 1993, 88 (01) : 332 - 332
  • [29] Automatic implantable cardioverter-defibrillators
    Smith, WM
    Ideker, RE
    [J]. ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 1999, 1 : 331 - 346
  • [30] Implantable cardioverter-defibrillators and driving
    Chen, Peng-Sheng
    [J]. HEART RHYTHM, 2007, 4 (05) : 688 - 689