THE CIRRUS IMMULITE AUTOMATED IMMUNOASSAY SYSTEM

被引:0
|
作者
BABSON, AL
机构
来源
JOURNAL OF CLINICAL IMMUNOASSAY | 1991年 / 14卷 / 02期
关键词
IMMUNOASSAY; AUTOMATION; CHEMILUMINESCENCE; RANDOM ACCESS;
D O I
暂无
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
The Cirrus Diagnostics IMMULITE(TM) is a random access automated immunoassay system built around a proprietary assay tube that provides for rapid and efficient washing of the captive antibody coated bead by spinning the tube on its vertical axis. Sample, excess reagent, and wash solution are captured in a coaxial waste sump integral with the tube. The alkaline phosphatase label is measured using a sensitive chemiluminescent substrate. System operation is very simple. The operator loads a carousel with label reagent for up to twelve resident assays. Samples in bar coded tubes are placed onto a continuous conveyer followed by up to five different assay tubes in any order. Samples can be added at any time and stats can be inserted as the next sample to be processed with no loss in throughput. Throughput is 120 tests per hour and the first result is available in 45 minutes. Periodic calibration of the stored standard curves is done with two calibrators which are loaded just like samples.
引用
收藏
页码:83 / 88
页数:6
相关论文
共 50 条
  • [1] Evaluation of an automated immunoassay method for cytokine measurement using the Immulite® immunoassay system
    Berthier, F
    Lambert, C
    Genin, C
    Bienvenu, J
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 1999, 37 (05) : 593 - 599
  • [2] Analytical and clinical performance of an automated immunoassay system (Immulite®) for estradiol in serum
    Rodríguez-Espinosa, J
    Otal-Entraigas, C
    Gascón-Roche, N
    Mora-Brugués, J
    Urgell-Rull, E
    Bordás-Serrat, JR
    Viscasillas-Molins, P
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 1998, 36 (12) : 969 - 974
  • [3] A RAPID AND AUTOMATED IMMUNOASSAY FOR CK-MB MEASUREMENT BY THE IMMULITE SYSTEM
    PANOSIANSAHAKIAN, N
    JACOBS, R
    LEI, JD
    SUSTARSIC, D
    ELSHAMI, AS
    [J]. CLINICAL CHEMISTRY, 1995, 41 (06) : S68 - S68
  • [4] An automated competitive immunoassay for androstenedione on the IMMULITE and IMMULITE 2000 analyzers.
    Wang, Z
    Bostanian, Z
    McKenna, NM
    Whitters, E
    Lei, JD
    Ito, T
    Hsia, P
    Panosian-Sahakian, N
    El Shami, AS
    [J]. CLINICAL CHEMISTRY, 2003, 49 (06) : A132 - A132
  • [5] MEASUREMENT OF DIGITOXIN, DIGOXIN AND THEOPHYLLINE IN SERUM BY THE IMMULITE AUTOMATED IMMUNOASSAY SYSTEM
    MILLER, S
    REES, J
    ELLIS, J
    PARKER, N
    BAINES, E
    KLINE, E
    ELSHAMI, AS
    [J]. CLINICAL CHEMISTRY, 1995, 41 (06) : S56 - S56
  • [6] An automated competitive immunoassay for vancomycin on the IMMULITE® 2000 and IMMULITE® 2500 analyzers.
    Wang, Z.
    Bostanian, Z.
    Zhong, D.
    Whitters, E.
    Lei, J.
    Panosian-Sahakian, N.
    Jaggi, K.
    Sustarsic, D.
    [J]. CLINICAL CHEMISTRY, 2006, 52 (06) : A77 - A77
  • [7] Laboratory evaluation of thyroid assays on the Immulite(R) automated immunoassay system.
    Le, Q
    Ka, M
    Chiu, L
    Toivola, B
    [J]. CLINICAL CHEMISTRY, 1996, 42 (06) : 253 - 253
  • [8] Automated chemiluminescent immunoassay of vancomycin on the IMMULITE® 2000 analyzer
    Bostanian, Z
    Panosian-Sahakian, N
    Chaturvedi, D
    Chaturvedi, M
    Whitters, E
    Lei, J
    Sustarsic, D
    El Shami, AS
    [J]. CLINICAL CHEMISTRY, 2004, 50 (06) : A131 - A131
  • [9] An automated immunoassay for gastrin on the IMMULITE 2000 analyzer.
    Whitters, E
    Lei, JD
    Figueroa, N
    Laroya, R
    Panosian-Sahakian, N
    El Shami, AS
    [J]. CLINICAL CHEMISTRY, 2003, 49 (06) : A131 - A131
  • [10] Automated chemiluminescent immunoassays for dsDNA antibody on the IMMULITE® 2000 and IMMULITE® 2500 immunoassay analyzers
    Qiu, B.
    Whitters, E.
    Lei, J.
    Thompson, S.
    Campbell, B.
    Jaggi, K.
    Sustarsic, D.
    [J]. CLINICAL CHEMISTRY, 2006, 52 (06) : A146 - A146