Effects of Eccentric Impeller Position on Radial Passive Stability in a Magnetically Levitated Centrifugal Blood Pump with a Double Volute

被引:8
|
作者
Shida, Shuya [1 ,3 ]
Masuzawa, Toru [2 ]
Osa, Masahiro [2 ]
Sato, Ryota [1 ]
机构
[1] Ibaraki Univ, Grad Sch Sci & Engn, Hitachi, Ibaraki, Japan
[2] Ibaraki Univ, Dept Mech Engn, Hitachi, Ibaraki, Japan
[3] 4-12-1 Nakanarusawa, Hitachi, Ibaraki 3168511, Japan
来源
关键词
eccentric impeller position; double volute; magnetic suspension; hydraulic force; computational fluid dynamics (CFD);
D O I
10.14326/abe.7.63
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Evaluation of the impeller radial stability is important from the bioengineering point of view in the development of mechanical circulatory support devices (MCSDs) for safer use as bridge for several months or destination therapy for years. In this study, radial stability of a magnetically levitated impeller in a centrifugal blood pump with an axially magnetic suspension system was evaluated by investigating the effects of the eccentric impeller position on passive stability, aiming to propose a pump design guide for the development and safer clinical use of MCSDs. First, impeller displacements in the prototype pump were measured using a mock loop together with laser displacement sensors. Then, the radial hydraulic forces exerted on an eccentric impeller were calculated using computational fluid dynamics (CFD) analysis for four volute-casing geometries. In addition, hemocompatibility was assessed using CFD calculations of scalar shear stress exerted on blood. Measurement of impeller displacement showed that the displacement varied from 0.56 to 0.27 mm at a rotational speed of 1800 rpm as the flow rate increased from 0 to 6.5 L/min. In the CFD calculation, the radial hydraulic force increased linearly from 0.2 to 1.7 N as the impeller displacement increased from 0 to 0.5 mm for all the double volute geometries, under conditions of a rotational speed of 1800 rpm and flow rates of 3, 5 and 7 L/min. These results indicate that the impeller stability in the prototype pump is acceptable at the operation conditions of ventricular assist devices, because the magnetic bearing stiffness of radial component was 4.1 N/mm. In the pressure recovery analysis of eccentric impellers, a double volute was not effective because of the unbalanced pressure field generated by the unbalanced pressure recovery. Thus, the increase in radial hydraulic force associated with an eccentric impeller could not be avoided by changing the conventional double volute design. The CFD analysis of geometrical variation indicated that widening of the radial clearance is an effective approach to improve the radial stability as well as hemocompatibility, although the radial clearance should be designed based on trade-offs among impeller stability, hemocompatibility and pump performance.
引用
收藏
页码:63 / 71
页数:9
相关论文
共 50 条
  • [1] Effects of Eccentric Impeller Position on Radial Passive Stability in a Magnetically Levitated Centrifugal Blood Pump with a Double Volute (vol 7, pg 63, 2018)
    Shida, Shuya
    Masuzawa, Toru
    Osa, Masahiro
    Sato, Ryota
    [J]. ADVANCED BIOMEDICAL ENGINEERING, 2018, 7 : 117 - 117
  • [2] Flow rate estimation of a centrifugal blood pump using the passively stabilized eccentric position of a magnetically levitated impeller
    Shidai, Shuya
    Masuzawa, Toru
    Osa, Masahiro
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2019, 42 (06): : 291 - 298
  • [3] Effects of gravity on flow rate estimations of a centrifugal blood pump using the eccentric position of a levitated impeller
    Shida, Shuya
    Masuzawa, Toru
    Osa, Masahiro
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2020, 43 (12): : 774 - 781
  • [4] Dynamic characteristics of a magnetically levitated impeller in a centrifugal blood pump
    Asama, Junichi
    Shinshi, Tadahiko
    Hoshi, Hideo
    Takatani, Setsuo
    Shimokohbe, Akira
    [J]. ARTIFICIAL ORGANS, 2007, 31 (04) : 301 - 311
  • [5] Hydraulic Analysis and Optimization in impeller of Magnetically Levitated Centrifugal Blood Pump
    Zhang, Huimin
    Zhu, Qi
    [J]. FRONTIERS OF MANUFACTURING AND DESIGN SCIENCE II, PTS 1-6, 2012, 121-126 : 1204 - 1208
  • [6] A compact highly efficient and low hemolytic centrifugal blood pump with a magnetically levitated impeller
    Asama, J
    Shinshi, T
    Hoshi, H
    Takatani, S
    Shimokohbe, A
    [J]. ARTIFICIAL ORGANS, 2006, 30 (03) : 160 - 167
  • [7] Mechanical antithrombogenic properties by vibrational excitation of the impeller in a magnetically levitated centrifugal blood pump
    Murashige, Tomotaka
    Hijikata, Wataru
    [J]. ARTIFICIAL ORGANS, 2019, 43 (09) : 849 - 859
  • [8] THE EFFECT OF IMPELLER POSITION ON CFD CALCULATIONS OF BLOOD FLOW IN MAGNETICALLY LEVITATED CENTRIFUGAL BLOOD PUMPS
    Fraser, Katharine H.
    Taskin, M. Ertan
    Zhang, Tao
    Richardson, J. Scott
    Gellman, Barry
    Dasse, Kurt
    Griffith, Bartley P.
    Wu, Zhongjun J.
    [J]. PROCEEDINGS OF THE ASME SUMMER BIOENGINEERING CONFERENCE, 2010, 2010, : 119 - 120
  • [9] Radial Passive Stability Analysis of Magnetically Levitated Impeller Position for the Development of Flow Rate Estimations of a Ventricular Assist Device
    Shida, Shuya
    Masuzawa, Toru
    Osa, Masahiro
    [J]. ADVANCED BIOMEDICAL ENGINEERING, 2022, 11 : 194 - 202
  • [10] CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER
    AKAMATSU, T
    NAKAZEKI, T
    ITOH, H
    [J]. ARTIFICIAL ORGANS, 1992, 16 (03) : 305 - 308