A BIOCOMPATIBLE NEEDLE-TYPE GLUCOSE SENSOR BASED ON PLATINUM-ELECTROPLATED CARBON ELECTRODE

被引:20
|
作者
CHEN, CY
TAMIYA, E
ISHIHARA, K
KOSUGI, Y
SU, YC
NAKABAYASHI, N
KARUBE, I
机构
[1] NATL TAIWAN UNIV,DEPT AGR CHEM,TAIPEI,TAIWAN
[2] ASAHI KENTO IND CO,MINAMI KU,YOKOHAMA,KANAGAWA,JAPAN
[3] TOKYO MED & DENT UNIV,INST MED & DENT ENGN,CHIYODA KU,TOKYO 101,JAPAN
关键词
BIOCOMPATIBILITY; 2-METHACRYLOYLOXYETHYL PHOSPHORYLCHOLINE (MPC); N-BUTYLMETHACRYLATE (BMA); CELLULOSE TRIACETATE (CTA); DIFFUSION-LIMITING EFFECT; GLUCOSE SENSOR; NEEDLE TYPE; GLUCOSE OXIDASE; SERUM;
D O I
10.1007/BF02921780
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
A biocompatible needle-type glucose sensor with a 3-electrode configuration was constructed. A platinum-electroplated carbon stick was used as the working electrode, Ag/AgCl as the reference electrode, and a disposable hypodermic needle made of stainless steel as the counter electrode. A Nation membrane, an immobilized glucose oxidase (GOD) membrane, and a biocompatible membrane with diffusion-limiting effect were coated successively onto the working electrode. The sensor showed a rapid response (< 120 s in batch operation), good reproducibility (RE < 3%), good stability (over 36 h in control serum), a wide dynamic range (5-600 mg/dL glucose), and superior biocompatibility. It was used to determine glucose in serum. The data obtained from the sensor showed good agreement with that from a clinical autoanalyzer (R > 0.95).
引用
收藏
页码:211 / 226
页数:16
相关论文
共 50 条
  • [1] BIOCOMPATIBLE NEEDLE-TYPE GLUCOSE SENSOR WITH POTENTIAL FOR USE IN-VIVO
    CHEN, CY
    SU, YC
    ISHIHARA, K
    NAKABAYASHI, N
    TAMIYA, E
    KARUBE, I
    [J]. ELECTROANALYSIS, 1993, 5 (04) : 269 - 276
  • [2] NEEDLE-TYPE GLUCOSE SENSOR
    SHICHIRI, M
    KAWAMORI, R
    YAMASAKI, Y
    [J]. METHODS IN ENZYMOLOGY, 1988, 137 : 326 - 334
  • [3] AMPEROMETRIC NEEDLE-TYPE GLUCOSE SENSOR BASED ON A MODIFIED PLATINUM-ELECTRODE WITH DIMINISHED RESPONSE TO INTERFERING MATERIALS
    CHEN, CY
    GOTOH, M
    MAKINO, H
    SU, YC
    TAMIYA, E
    KARUBE, I
    [J]. ANALYTICA CHIMICA ACTA, 1992, 265 (01) : 5 - 14
  • [4] Carbon nanotubes based glucose needle-type biosensor
    Jia, Jinyan
    Guan, Wenjun
    Sim, Minghao
    Li, Yongquan
    Li, Hong
    [J]. SENSORS, 2008, 8 (03): : 1712 - 1718
  • [5] Development of a needle-type glucose sensor based on a titanium dioxide oxygen electrode for the artificial pancreas
    Ikeda, Shoichiro
    Hara, Masanori
    Hishida, Yuko
    Kimura, Masami
    Ito, Kaname
    [J]. Sensors and Actuators, B: Chemical, 1993, B13 (1 -3 pt 1) : 315 - 318
  • [6] Ferrocene-mediated needle-type glucose sensor covered with newly designed biocompatible membrane
    Sakakida, Michiharu
    Nishida, Kenro
    Shichiri, Motoaki
    Ishihara, Kazuhiko
    Nakabayashi, Nobuo
    [J]. Sensors and Actuators, B: Chemical, 1993, B13 (1 -3 pt 1) : 319 - 322
  • [7] FERROCENE-MEDIATED NEEDLE-TYPE GLUCOSE SENSOR COVERED WITH NEWLY DESIGNED BIOCOMPATIBLE MEMBRANE
    SAKAKIDA, M
    NISHIDA, K
    SHICHIRI, M
    ISHIHARA, K
    NAKABAYASHI, N
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1993, 13 (1-3) : 319 - 322
  • [8] DEVELOPMENT OF A NOUVELLE NEEDLE-TYPE GLUCOSE SENSOR BASED ON TITANIUM-OXIDE OXYGEN-ELECTRODE
    IKEDA, S
    KIMURA, M
    NODA, H
    ITO, K
    OHKURA, K
    NAKAO, A
    TAKAGI, H
    KONDO, T
    [J]. DENKI KAGAKU, 1988, 56 (12): : 1055 - 1062
  • [9] Development of needle-type glucose sensor with high selectivity
    Yang, QL
    Atanasov, P
    Wilkins, E
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1998, 46 (03) : 249 - 256
  • [10] Preparation and characterization of electrochemically immobilized needle-type glucose sensor
    Cao, DJ
    Zhou, M
    Wu, XQ
    [J]. DENKI KAGAKU, 1996, 64 (12): : 1274 - 1275