A DEVICE FOR INVITRO IRRADIATION WITH ALPHA-PARTICLES USING AN ALPHA-EMITTING RADIOACTIVE SOURCE

被引:3
|
作者
ISHIGURE, N
NAKANO, T
ENOMOTO, H
机构
[1] Division of Comparative Radiotoxicology, National Institute of Radiological Sciences, 4-9-1 Anagawa, Chiba
关键词
DEVICE; INVITRO; IRRADIATION; ALPHA-PARTICLE; ALPHA-SOURCE;
D O I
10.1269/jrr.32.404
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
A device to irradiate a monolayer of cultured celts with alpha-particles using an Am-241 alpha-source (33.4 MBq) was designed to investigate RBEs of alpha-particles in cell killing, induction of chromosome aberration, mutagenic changes and transformation. This device can be used conveniently in a common laboratory by a small number of researchers without any limitation of machine time. The device performs as follows: (1) The energy of alpha-particles at the entrance of the cell layer is 3.20 MeV with a standard deviation of 0.25 MeV, (2) the incident angle to the cell layer is 82.8 degrees with a standard deviation of 3.2 degrees, (3) the fluence rate is 4.7 x 10(5) cm-2.min-1, (4) the average LET(infinity) for a cell layer 5-mu-m thick is 138 keV/mu-m, (5) the average dose rate for a cell layer 5-mu-m thick is 0.10 Gy/min., (6) a temperature and CO2 concentration conducive to cell cultivation are maintained during irradiation.