DOES THE PHOTON-BEAM QUALITY FACTOR DEPEND ON THE TYPE OF IONIZATION-CHAMBER

被引:2
|
作者
CHRIST, G
MAJOR, J
NUSSLIN, F
机构
[1] Abteilung Medizinische Physik, Radiologische Universitätsklinik Tübingen, Hoppe-Seylerstr. 3
关键词
PHOTON BEAM QUALITY FACTOR; TPR(10)(20); IONIZATION DOSIMETRY;
D O I
10.1016/0167-8140(95)01602-D
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The photon beam quality factor Q of 6- and 15-MV photon beams has been measured using different types of thimble ionization chambers with various wall, electrode and insulator materials. Some types of chambers seem to overestimate the corresponding e-value by 0.4% (6 MV) and 0.7%(15 MV), respectively. The effect on the stopping power ratio water to air and on the pertubation correction will be discussed.
引用
收藏
页码:153 / 155
页数:3
相关论文
共 50 条
  • [31] THE PERFORMANCE OF A PLATE-TYPE PHOTON-BEAM MONITOR AT SRRC
    CHANG, CN
    CHEN, CI
    LEE, SB
    TSENG, PC
    HSU, KT
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 291 (1-2): : 419 - 421
  • [32] COLLECTION EFFICIENCY OF AN IONIZATION-CHAMBER IN A PULSED SWEPT BEAM - COLLIMATOR SCATTER EFFECTS
    WEINHOUS, MS
    MELI, JA
    PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (10): : 1147 - 1155
  • [33] A new reference-type ionization chamber with direction-independent response for use in small-field photon-beam dosimetry - An experimental and Monte Carlo study
    Delfs, Bjoern
    Kapsch, Ralf-Peter
    Chofor, Ndimofor
    Looe, Hui Khee
    Harder, Dietrich
    Poppe, Bjoern
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2019, 29 (01): : 39 - 48
  • [34] A NEW LOOK AT DISPLACEMENT FACTOR AND POINT OF MEASUREMENT CORRECTIONS IN IONIZATION-CHAMBER DOSIMETRY
    AWSCHALOM, M
    ROSENBERG, I
    TENHAKEN, RK
    MEDICAL PHYSICS, 1983, 10 (03) : 307 - 313
  • [35] A NEW TYPE HYGROMETER USING A DOUBLE IONIZATION-CHAMBER WITH ALPHA-RAYS
    MATSUMOTO, S
    KOBAYASHI, H
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 255 (1-2): : 290 - 292
  • [36] DETERMINATION OF MEGAVOLTAGE PHOTON-BEAM QUALITY USING THE DOSE PERTURBATION AT INTERFACES
    DAS, IJ
    KHAN, FM
    GERBI, BJ
    MEDICAL PHYSICS, 1987, 14 (03) : 491 - 491
  • [37] TEST OF A DUAL-TYPE GRIDDED IONIZATION-CHAMBER USING LIQUID XENON
    MASUDA, K
    HITACHI, A
    HOSHI, Y
    DOKE, T
    NAKAMOTO, A
    SHIBAMURA, E
    TAKAHASHI, T
    NUCLEAR INSTRUMENTS & METHODS, 1980, 174 (03): : 439 - 446
  • [38] A PRELIMINARY COMPARISON OF LYOLUMINESCENCE DOSIMETRY OF A PI- MESON BEAM WITH IONIZATION-CHAMBER MEASUREMENTS
    BARTLETT, DT
    ETTINGER, KV
    READING, DH
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1981, 32 (08): : 603 - 605
  • [39] IONIZATION-CHAMBER TYPE OF FOCAL-PLANE DETECTOR FOR HEAVY-IONS
    ERSKINE, JR
    BRAID, TH
    STOLTZFUS, JC
    NUCLEAR INSTRUMENTS & METHODS, 1976, 135 (01): : 67 - 82
  • [40] Ionization chamber radial response deconvolution in megavoltage photon beam
    Kulmala, A.
    Tenhunen, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (18): : 7505 - 7519