DOES THE PHOTON-BEAM QUALITY FACTOR DEPEND ON THE TYPE OF IONIZATION-CHAMBER

被引:2
|
作者
CHRIST, G
MAJOR, J
NUSSLIN, F
机构
[1] Abteilung Medizinische Physik, Radiologische Universitätsklinik Tübingen, Hoppe-Seylerstr. 3
关键词
PHOTON BEAM QUALITY FACTOR; TPR(10)(20); IONIZATION DOSIMETRY;
D O I
10.1016/0167-8140(95)01602-D
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The photon beam quality factor Q of 6- and 15-MV photon beams has been measured using different types of thimble ionization chambers with various wall, electrode and insulator materials. Some types of chambers seem to overestimate the corresponding e-value by 0.4% (6 MV) and 0.7%(15 MV), respectively. The effect on the stopping power ratio water to air and on the pertubation correction will be discussed.
引用
收藏
页码:153 / 155
页数:3
相关论文
共 50 条
  • [1] Does the photon beam quality factor depend on the type of ionization chamber? (vol 36, pg 153, 1995)
    Christ, G
    Major, J
    Nusslin, F
    RADIOTHERAPY AND ONCOLOGY, 1996, 38 (01) : 81 - 81
  • [2] IONIZATION-CHAMBER MEASUREMENTS OF A PION BEAM
    SHORTT, KR
    HENKELMAN, RM
    RADIATION RESEARCH, 1981, 85 (03) : 419 - 431
  • [3] A MULTIWIRE CHAMBER FOR POSITRON AND PHOTON-BEAM SCANNING
    ALBICOCCO, M
    CAPITANI, GP
    DESANCTIS, E
    DIGIACOMO, P
    GUARALDO, C
    LUCHERINI, V
    POLLI, E
    REOLON, AR
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 203 (1-3): : 63 - 68
  • [4] A NEW TYPE OF IONIZATION-CHAMBER TELESCOPE
    LI, SL
    HU, XQ
    ZHU, YT
    WANG, XM
    WANG, GS
    XIE, YX
    SHEN, WQ
    CHINESE PHYSICS, 1989, 9 (04): : 1053 - 1057
  • [5] BEAM QUALITY SPECIFICATION FOR PHOTON-BEAM DOSIMETRY
    KOSUNEN, A
    ROGERS, DWO
    MEDICAL PHYSICS, 1993, 20 (04) : 1181 - 1188
  • [6] SMALL IONIZATION-CHAMBER FOR MEASUREMENT OF NEUTRON AND PHOTON DOSE
    SPOKAS, JJ
    WILLIAMSON, FS
    HOLMBLAD, GL
    RADIATION RESEARCH, 1974, 59 (01) : 53 - 54
  • [7] MEASURING THE INTENSITY OF THE PHOTON-BEAM ON A SETUP WITH A STREAMER CHAMBER
    ARKATOV, YM
    VATSET, PI
    VOLOSHCHUK, VI
    ZOLENKO, VA
    PROKHORETS, IM
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1981, 24 (01) : 39 - 42
  • [8] EXTERNAL BEAM PROFILES MEASURED WITH FILM OR IONIZATION-CHAMBER
    LAM, WC
    LAM, KS
    ONEILL, MJ
    PARTOWMAH, M
    KHALIFEH, A
    MEDICAL PHYSICS, 1985, 12 (04) : 540 - 540
  • [9] EVALUATION OF AVERAGE QUALITY FACTOR IN A MIXED RADIATION-FIELD WITH AN IONIZATION-CHAMBER
    ODA, K
    YAMAMOTO, T
    KAWANISHI, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 202 (03): : 481 - 486
  • [10] THEORY OF MEASUREMENT OF COLLISION KERMA IN PHOTON BEAMS WITH AN IONIZATION-CHAMBER
    SHIRAGAI, A
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1990, 29 (06): : 1227 - 1229