The tympanic thermometer in pediatrics as an alternative to the mercury-in-glass thermometer

被引:4
|
作者
Victoria Vertedor-Hurtado, Maria [1 ]
Padin-Lopez, Susana [1 ]
Jose Carreira-Pastor, Maria [1 ]
Maria Lopez-Martinez, Julia [1 ]
机构
[1] Hosp Materno Infantil, HRU Carlos Haya, Urgencias Pediat, Malaga, Spain
来源
ENFERMERIA CLINICA | 2009年 / 19卷 / 03期
关键词
Pediatrics; Fever of unknown origin; Body temperature; Triage;
D O I
10.1016/j.enfcli.2008.10.016
中图分类号
R47 [护理学];
学科分类号
1011 ;
摘要
Objectives: To determine the behavior of a new and smaller model of infrared thermometer cone and to assess whether it is an appropriate alternative to determine fever in children under 14-years-old. Method: We performed a cross-sectional, descriptive study comparing the temperatures taken with a mercury thermometer with those taken by an infrared thermometer in children under 14-years-old. The researchers followed the manufacturer's instructions for use for each thermometer. The overall sample of 400 children was divided into two age groups. Group I consisted of children aged less than 2 years and group II of 2-14-years-old. In group I the rectal mercury thermometer was placed in the rectum for 3 minutes and the tympanic thermometer was used in rectal mode and placed in the right ear. In group II, the axillary mercury thermometer was placed for 8 minutes in the right axilla and the tympanic thermometer was used in the axillary mode. To analyze intraobserver bias, 50 patients were selected from the sample and their temperatures were simultaneously taken by two nurses from the team. The intraclass correlation coefficient (ICC) was used both to measure the reliability of the tympanic thermometer and to analyze intraobserver bias. Results: Temperature measurements with both instruments showed an ICC of 0.91 (95% confidence interval [CI], 0.88-0.94) for group I and 0.90 (95% CI, 0.87-0.92) for group II. The reproducibility of the measurements taken by the two nurses in 50 patients showed an ICC of 0.97 (95% CI, 0.95-0.98) for the tympanic thermometer. Conclusions: The infrared thermometer is an appropriate device for rapidly measuring temperature in the emergency department. However, the measurements taken should be confirmed by another method when clinical decisions are based on temperature values. (C) 2008 Elsevier Espana, S.L. All rights reserved.
引用
收藏
页码:115 / 120
页数:6
相关论文
共 50 条
  • [41] MERCURY SPILLED FROM A BROKEN THERMOMETER
    HENSCHLER, D
    [J]. DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1983, 108 (10) : 395 - 395
  • [42] MERCURY EXPOSURE AND EFFECTS AT A THERMOMETER FACTORY
    RICHTER, ED
    PELED, N
    LURIA, M
    [J]. SCANDINAVIAN JOURNAL OF WORK ENVIRONMENT & HEALTH, 1982, 8 : 161 - 166
  • [43] SHOULD WE BAN THE MERCURY THERMOMETER
    STUART, C
    [J]. JOURNAL OF THE ROYAL SOCIETY OF MEDICINE, 1993, 86 (02) : 122 - 122
  • [45] A non-mercury thermometer alternative for use in older melting point apparatuses
    Ongley, Lois K.
    Kern, Clayton S.
    Woods, Barry W.
    [J]. JOURNAL OF CHEMICAL EDUCATION, 2008, 85 (09) : 1263 - 1264
  • [46] Unreliability of the infrared tympanic thermometer in clinical practice: A comparative study with oral mercury and oral electronic thermometers
    Modell, JG
    Katholi, CR
    Kumaramangalam, SM
    Hudson, EC
    Graham, D
    [J]. SOUTHERN MEDICAL JOURNAL, 1998, 91 (07) : 649 - 654
  • [47] Does central tympanic membrane perforation affect infrared tympanic thermometer measurements in adults?
    Hamdi Tasli
    Mert Cemal G?kg?z
    [J]. Journal of Otology, 2018, 13 (04) : 128 - 130
  • [48] Comparison of tympanic temperature readings with axillary thermometer in the OPD setting
    Malik, Munir
    Rabbani, Mohammad Zafar
    Amir, Mohammed
    Afzal, Uzma
    Malik, Mohammad Umair
    [J]. RAWAL MEDICAL JOURNAL, 2009, 34 (02):
  • [49] Thermometer glass at higher temperatures.
    McClellan, W
    [J]. JOURNAL OF THE FRANKLIN INSTITUTE, 1901, 152 : 63 - 72
  • [50] ELECTRONIC LIQUID-IN-GLASS THERMOMETER
    SAARIMAA, R
    WALLIN, P
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1976, 47 (02): : 195 - 197