DETERMINATION OF URANIUM AND THORIUM IN ALUMINUM WITH FLOW-INJECTION AND LASER ABLATION INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY

被引:13
|
作者
VANDEWEIJER, P
VULLINGS, PJMG
BAETEN, WLM
DELAAT, WJM
机构
[1] Philips Research Laboratories, 5600 JA Eindhoven
关键词
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY; ORIFICE DEPOSITION; FLOW INJECTION; LASER ABLATION; SPIKE PHENOMENON;
D O I
10.1039/ja9910600609
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In order to determine uranium and thorium at the sub-ng g-1 level in aluminium, the limit of detection (LOD) for continuous-flow nebulization inductively coupled plasma mass spectrometry (ICP-MS) is not sufficient, when a sample solution with the usual maximum concentration of 1 mg ml-1 is used. Therefore, two alternative sample introduction techniques have been used, flow injection (FI) and laser ablation (LA). With FI-ICP-MS the achievement of sub-ng g-1 detection limits is hampered by the presence of 'spikes'. Although these spikes are also present with LA, it is possible to obtain a 0.2 ng g-1 LOD for uranium and thorium. This LOD is achieved artificially, by rejecting all measurements containing spikes.