A Monte Carlo simulation of coincidence detection and imaging of gamma-ray cascades with a scintillation camera

被引:4
|
作者
Pahlka, R. Benton [1 ]
Kappadath, S. Cheenu [1 ]
Mawlawi, Osama R. [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Imaging Phys, UNIT 1472, POB 301402, Houston, TX 77230 USA
来源
关键词
gamma-ray cascades; coincidence detection; gamma camera;
D O I
10.1088/2057-1976/aad572
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Several radionuclei used in nuclear medicine decay through the emission of cascaded gamma-rays. These successively-emitted gamma-rays could potentially be detected in coincidence and then be used to create coincident event-based images. While several groups have considered various ways to leverage the gamma-ray cascade to produce images, the concept of producing images using cascaded gamma rays in coincidence with a dual-head gamma camera has not been previously explored in detail. Given that current dual-head gamma cameras cannot be configured in coincidence mode, the purpose of this work was to explore this concept with In-111 using a Monte Carlo simulation. First, GEANT4 was used to model a dual-head gamma camera and to simulate the decay of the nucleus. The model was then validated using planar data from a physical scanner. The validated model was then used to simulate a point source of In-111 located at the camera isocenter to evaluate the true coincidence efficiency with the detectors oriented at 90 degrees. The noise-equivalent-count-rate (NECR) as a function of timing window resolution was computed for different source activities. A simple reconstruction algorithm was then used to produce coincidence-based images from sources in various configurations. The coincident gamma-ray detection efficiency for a single point source in air with the detector heads oriented at 90 degrees was found to be 6 x 10(-8), independent of the source-to-detector distance. The time window resolution for the maximum NECR ranged between 200 and 500 ns, dependent on source activity. The algorithm could produce images over a range of timing window resolutions, with varying levels of reconstruction artifacts. Images produced from a small acrylic phantom showed that scattered events produced small degradations in resolution and contrast-to-noise ratios. This work described the conditions under which images based on coincident events from In-111 could be created using a dual head gamma camera, assuming appropriate electronics modifications.
引用
收藏
页数:18
相关论文
共 50 条
  • [11] Monte Carlo Simulation of the TAIGA Hybrid Gamma-Ray Experiment
    A. Grinyuk
    E. Postnikov
    L. Sveshnikova
    Physics of Atomic Nuclei, 2020, 83 : 262 - 267
  • [12] Monte Carlo Simulation of the TAIGA Hybrid Gamma-Ray Experiment
    Grinyuk, A.
    Postnikov, E.
    Sveshnikova, L.
    PHYSICS OF ATOMIC NUCLEI, 2020, 83 (02) : 262 - 267
  • [13] MONTE-CARLO SIMULATION OF ATMOSPHERIC GAMMA-RAY PRODUCTION
    MORRIS, DJ
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1980, 25 (04): : 528 - 528
  • [14] MONTE-CARLO SIMULATION OF PLUTONIUM GAMMA-RAY STANDARDS
    WANG, TF
    CARLSON, JB
    KOENIG, ZM
    RUHTER, WD
    LEE, TSH
    WINN, J
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 353 (1-3): : 678 - 682
  • [15] MONTE-CARLO SIMULATION OF A NEW GAMMA-RAY TELESCOPE
    SIMONE, J
    ONEILL, T
    TUMER, OT
    ZYCH, AD
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1985, 32 (01) : 124 - 128
  • [16] MONTE-CARLO SIMULATION OF ATMOSPHERIC GAMMA-RAY SCATTERING
    MORRIS, DJ
    AIP CONFERENCE PROCEEDINGS, 1984, (115) : 665 - 668
  • [17] Monte Carlo modeling of neutron and gamma-ray imaging systems
    Hall, J
    NEUTRONS IN RESEARCH AND INDUSTRY, INTERNATIONAL CONFERENCE, 1997, 2867 : 465 - 468
  • [18] MEASUREMENT OF LOW GAMMA-RAY WITH THE SCINTILLATION CAMERA
    KLOSE, E
    RITTER, EP
    ENDERT, G
    RADIOBIOLOGIA-RADIOTHERAPIA, 1979, 20 (06) : 826 - 826
  • [19] A gamma-ray imaging camera for ambient radioactivity detection
    Altomare, C.
    Di Venere, L.
    Fanchini, E.
    Giordano, F.
    Loparco, F.
    Morichi, M.
    Spinelli, P.
    Swiderski, L.
    2019 IEEE 8TH INTERNATIONAL WORKSHOP ON ADVANCES IN SENSORS AND INTERFACES (IWASI), 2019, : 297 - 301
  • [20] A gamma-ray imaging camera for ambient radioactivity detection
    Altomare, C.
    Di Venere, L.
    Fanchini, E.
    Giordano, F.
    Loparco, F.
    Morichi, M.
    Pantaleo, F. R.
    Spinelli, P.
    Swiderski, L.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 981