Using the Monte Carlo method, we calculated the limit to the intrinsic spatial resolution of scintillation gamma-cameras, which is determined by gamma-quanta interacting with the scintillator. The calculations were for a scintillator in the shape of an infinitely large flat uniform layer of finite thickness, onto which an infinitely narrow beam of gamma-quanta is normally incident. The intrinsic spatial resolution was calculated for NaI(Tl), CsI(Tl), and Bi4Ge3O12 scintillators with thicknesses varying from (1/4) to 3 in. and for incident gamma-radiation energies ranging in between 0.5-5 MeV.
引用
收藏
页码:617 / 621
页数:5
相关论文
共 50 条
  • [1] THE NOISE-EQUIVALENT BANDPASS AS A MEASURE OF SPATIAL-RESOLUTION FOR GAMMA-CAMERAS
    SHOSA, DW
    KAUFMAN, L
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1980, 21 (06) : P81 - P82
  • [2] SPATIAL-RESOLUTION OF LARGE FIELD OF VIEW GAMMA-CAMERAS FOR TL-201 MYOCARDIAL IMAGING
    SARPER, R
    FAJMAN, WA
    WITKOWSKI, J
    TARCAN, YA
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1978, 19 (06) : 731 - 731
  • [3] TWO-DIMENSIONAL SPATIAL DISTORTION CORRECTION ALGORITHM FOR SCINTILLATION GAMMA-CAMERAS
    JIH, F
    KING, SE
    LIM, CB
    CHANEY, R
    GRAY, E
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1985, 26 (05) : P101 - P101
  • [4] Extrinsic Versus Intrinsic Uniformity Correction for gamma-Cameras
    Bolstad, Randy
    Brown, Jody
    Grantham, Vesper
    [J]. JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY, 2011, 39 (03) : 208 - 212
  • [5] SPATIAL-RESOLUTION OF THERMOGRAPHIC CAMERAS
    STEKETEE, J
    [J]. BIBLIOTHECA RADIOLOGICA, 1975, (06) : 25 - 32
  • [6] The variation of intrinsic spatial resolution across the UFOV of scintillation cameras
    Papanastassiou, Emmanouil K.
    Psarrakos, Kyriakos
    Sioundas, Anastasios
    Ballas, Apostolos
    Koufogiannis, Dimitrios
    Hatziioannou, Konstantinos
    [J]. COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2006, 30 (08) : 417 - 426
  • [7] RESOLUTION OF PINHOLE COLLIMATORS USED WITH GAMMA-CAMERAS
    GOULET, P
    JOY, MLG
    [J]. MEDICAL PHYSICS, 1980, 7 (05) : 571 - 573
  • [8] DATA LOSS AND COUNT RATE POTENTIAL IN SCINTILLATION GAMMA-CAMERAS
    KING, SE
    LIM, CB
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1985, 26 (05) : P101 - P102
  • [9] SPATIAL DISTORTION CORRECTION METHOD FOR GAMMA-CAMERAS
    STOUB, EW
    COLSHER, JG
    LEE, PC
    MUEHLLEHNER, G
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1979, 20 (06) : 608 - 608
  • [10] PERFORMANCE EVALUATIONS OF RECENT WIDE FIELD SCINTILLATION GAMMA-CAMERAS
    LIM, CB
    HOFFER, PB
    ROLLO, FD
    LILIEN, DL
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1978, 19 (08) : 942 - 947