CALIBRATION STANDARD FOR LASER-DOPPLER PERFUSION MEASUREMENTS

被引:20
|
作者
LIEBERT, A [1 ]
LEAHY, M [1 ]
MANIEWSKI, R [1 ]
机构
[1] UNIV OXFORD,CHURCHILL HOSP,DEPT DERMATOL,OXFORD OX3 7LJ,ENGLAND
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 1995年 / 66卷 / 11期
关键词
D O I
10.1063/1.1146145
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The objective of the present study was the development of a new calibration standard for the laser-Doppler perfusion monitors. The beat frequency spectrum obtained from a solution of latex spheres undergoing Brownian motion was compared to that obtained from a porous polyethylene material containing the same solution. The mixture of static and dynamic scattering events in the latter case was expected to produce a spectrum similar to that obtained in the real human tissue. The results obtained demonstrate clearly the difference between the signals recorded from the commonly used motility standard which are the result of multiple Doppler shifting and homodyne mixing, and those generated from our enhanced model which predominantly result from single Doppler shifting and heterodyne mixing. (C) 1995 American Institute of Physics.
引用
收藏
页码:5169 / 5173
页数:5
相关论文
共 50 条
  • [1] LASER-DOPPLER MEASUREMENTS OF PERIPHERAL PERFUSION IN HYPOVOLEMIC SHOCK PATIENTS
    SCHELONKA, LP
    CLEVENGER, FW
    CLINICAL RESEARCH, 1994, 42 (01): : A9 - A9
  • [2] LASER-DOPPLER FLOWMETRY - A NONINVASIVE METHOD FOR MICROVASCULAR PERFUSION MEASUREMENTS
    OBERG, PA
    INTERNATIONAL JOURNAL OF MICROCIRCULATION-CLINICAL AND EXPERIMENTAL, 1986, 5 (2-3): : 163 - 163
  • [3] BIOLOGICAL ZERO IN LASER-DOPPLER PERFUSION MEASUREMENTS - STUDIES IN NEONATES
    NORMAN, M
    MARTIN, H
    HERIN, P
    ACTA PAEDIATRICA, 1995, 84 (04) : 457 - 459
  • [4] Multichannel laser-Doppler probe for blood perfusion measurements with depth discrimination
    Liebert, A.
    Leahy, M.
    Maniewski, R.
    Medical and Biological Engineering and Computing, 1998, 36 (06): : 740 - 747
  • [5] Evaluation of different signal processing algorithms in Laser-Doppler perfusion measurements
    Leahy, M
    Liebert, A
    Maniewski, R
    OPTO-IRELAND 2002: OPTICS AND PHOTONICS TECHNOLOGIES AND APPLICATIONS, PTS 1 AND 2, 2003, 4876 : 120 - 127
  • [6] Multichannel laser-Doppler probe for blood perfusion measurements with depth discrimination
    Liebert, A
    Leahy, M
    Maniewski, R
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1998, 36 (06) : 740 - 747
  • [7] Multichannel laser-Doppler probe for blood perfusion measurements with depth discrimination
    A. Liebert
    M. Leahy
    R. Maniewski
    Medical and Biological Engineering and Computing, 1998, 36 : 740 - 747
  • [8] PHOTODETECTORS FOR LASER-DOPPLER MEASUREMENTS
    DURST, F
    SENDER, J
    EXPERIMENTS IN FLUIDS, 1990, 9 (1-2) : 13 - 16
  • [9] Photodetectors for laser-Doppler measurements
    Durst, F.
    Sender, J.
    Experiments in Fluids, 1990, 9 (1-2): : 13 - 16
  • [10] Influence of probe optical arrangement on biological zero in laser-Doppler perfusion measurements
    Liebert, A
    Maniewski, R
    PROCEEDINGS OF THE 18TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 18, PTS 1-5, 1997, 18 : 200 - 201