APPLICATION OF AN AUTOMATED PLASMA FILTRATION DEVICE AND A BLOOD MONITORING-SYSTEM FOR LDL APHERESIS

被引:0
|
作者
TABORSKI, U [1 ]
KOHLER, T [1 ]
SCHUPPHAUS, S [1 ]
HOFMANN, MR [1 ]
MULLERBERGHAUS, G [1 ]
机构
[1] KERCKHOFF KLIN,DEPT HEMOSTASEOL & TRANSFUS MED,W-6350 BAD NAUHEIM,GERMANY
来源
关键词
AUTOMATED PLASMAPHERESIS; PLASMA FILTRATION; PLASMA SEPARATION; IMMUNOGLOBULINS; BLOOD CELL COUNTS; LDL APHERESIS;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study describes the adaptation of a filtration separation device (AK 10 from Gambro, Munchen, FRG) and a blood monitor system (BTS 100 from Diamed, Koln, FRG) to the automated adsorption/desorption device for LDL apheresis. Plasma separation by filtration was performed by a hollow-fiber filter (OP5, Asahi Corp., Japan). In order to evaluate the quality of the plasma obtained by filtration, the platelet contamination, the activity of factors V and VIII and the concentration of the immunoglobulins IgG and IgM were determined. The following results were found: Platelet contamination of the plasma was 2,740/mul. As compared to the patient's pretreatment values, activity of factor V was 94%, activity of factor VIII was 104%, concentration of IgG was 96%, concentration of IgM was 94%. Ten LDL aphereses were evaluated: red cell counts, platelet counts, hemoglobin concentration and hematocrit were not changed significantly. Thus, this new combination system represents an improved alternative to previously performed centrifugation methods.