INFLUENCE OF INFUSION-PUMP OPERATION AND FLOW-RATE ON HEMODYNAMIC STABILITY DURING EPINEPHRINE INFUSION

被引:42
|
作者
KLEM, SA
FARRINGTON, JM
LEFF, RD
机构
[1] UNIV KANSAS,MED CTR,DEPT PEDIAT,KANSAS CITY,KS 66103
[2] UNIV KANSAS,MED CTR,DEPT PHARM,KANSAS CITY,KS 66103
关键词
INFUSION PUMPS; EPINEPHRINE; METHOHEXITAL; BLOOD PRESSURE; DRUG DELIVERY SYSTEMS; INFUSIONS; INTRAVENOUS; HEMODYNAMICS; CATECHOLAMINES; CARDIAC EMERGENCIES; CRITICAL ILLNESS;
D O I
10.1097/00003246-199308000-00023
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To determine whether variations in the flow rate of epinephrine solutions administered via commonly available infusion pumps lead to significant variations in blood pressure (BP) in vivo. Design: Prospective, randomized, crossover study with factorial design, using infusion pumps with four different operating mechanisms (pulsatile diaphragm, linear piston/syringe, cyclic piston-valve, and linear peristaltic) and three drug delivery rates (1, 5, and 10 mL/hr). Subjects: Two healthy, mixed-breed dogs (12 to 16 kg). Interventions: Dogs were made hypotensive with methohexital bolus and continuous infusion. BP was restored to normal with constant-dose epinephrine infusion via two pumps at each rate. Measurements: Femoral mean arterial pressure (MAP) was recorded every 10 secs. Pump-flow continuity was quantitated in vitro using a digital gravimetric technique. Variations in MAP and flow continuity were expressed by the coefficient of variation; analysis of variance was used for comparisons. Results: The mean coefficients of variations for MAP varied from 3.8 +/- 3.1% (linear piston/syringe) to 6.1 +/- 6.6% (linear peristaltic), and from 3.4 +/- 2.2% (10 mL/hr) to 7.9 +/- 6.6% (1 mL/hr). The coefficients of variation for in vitro flow continuity ranged from 9 +/- 8% (linear piston-syringe) to 250 +/- 162% (pulsatile diaphragm), and from 35 +/- 44% (10 mL/hr) to 138 +/- 196% (1 mL/hr). Both the type of pump and infusion rate significantly (p < .001) influenced variation in drug delivery rate. The 1 mL/hr infusion rate significantly (p <.01) influenced MAP variation. Cyclic fluctuations in MAP of less-than-or-equal-to 30 mm Hg were observed using the pulsatile diaphragm pump at 1 mL/hr. Conclusion: Factors inherent in the operating mechanisms of infusion pumps may result in clinically important hemodynamic fluctuations when administering a concentrated short-acting vasoactive medication at slow infusion rates.
引用
收藏
页码:1213 / 1217
页数:5
相关论文
共 50 条
  • [41] Hemodynamic response to intentionally altered flow continuity of dobutamine and dopamine by an infusion pump in infants
    Stowe, CD
    Storgion, SA
    Lee, KR
    Phelps, SJ
    [J]. PHARMACOTHERAPY, 1996, 16 (06): : 1018 - 1023
  • [42] FACTORS AFFECTING DIAZEPAM INFUSION - SOLUBILITY, ADMINISTRATION-SET COMPOSITION, AND FLOW-RATE
    MASON, NA
    CLINE, S
    HYNECK, ML
    BERARDI, RR
    HO, NFH
    FLYNN, GL
    [J]. AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1981, 38 (10): : 1449 - 1454
  • [43] Disturbance of Vancomycin Infusion Flow during Multidrug Infusion: Influence on Endothelial Cell Toxicity
    Drouet, Maryline
    Cuvelier, Elodie
    Chai, Feng
    Genay, Stephanie
    Odou, Pascal
    Decaudin, Bertrand
    [J]. ANTIBIOTICS-BASEL, 2022, 11 (01):
  • [44] In-use stability of Rituximab and IVIG during intravenous infusion: Impact of peristaltic pump, IV bags, flow rate, and plastic syringes
    Hada, Shavron
    Shin, I. Jeong
    Park, Ha Eun
    Kim, Ki Hyun
    Kim, Kwang Joon
    Jeong, Seong Hoon
    Kim, Nam Ah
    [J]. INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2024, 663
  • [45] Performance of a linear peristaltic infusion pump during red blood cells administration and the influence of infusion rates
    DM Kusahara
    M Pedreira
    AM Avelar
    MA Peterlini
    [J]. Intensive Care Medicine Experimental, 3 (Suppl 1)
  • [46] EFFECTS OF EPINEPHRINE AND NOREPINEPHRINE ON HEMODYNAMIC PARAMETERS AND ARRHYTHMIAS DURING A CONTINUOUS-INFUSION OF AMITRIPTYLINE IN RATS
    KNUDSEN, K
    ABRAHAMSSON, J
    [J]. JOURNAL OF TOXICOLOGY-CLINICAL TOXICOLOGY, 1993, 31 (03): : 461 - 471
  • [47] FLOW DYNAMICS IN INDIVIDUAL NEPHRONS OF RAT KIDNEY DURING INTRAVENOUS INFUSION OF EPINEPHRINE
    MANGOS, JA
    BRAUN, G
    [J]. PFLUGERS ARCHIV FUR DIE GESAMTE PHYSIOLOGIE DES MENSCHEN UND DER TIERE, 1966, 289 (02): : R79 - &
  • [48] Effects of mental workload on nurses' visual behaviors during infusion pump operation
    Kataoka, Jun
    Sasaki, Minako
    Kanda, Katsuya
    [J]. JAPAN JOURNAL OF NURSING SCIENCE, 2011, 8 (01) : 47 - 56
  • [49] Influence of temperature on the flow rate of Baxter™ elastomeric infusion pumps
    Ehrmann, S.
    Delpech, M.
    Fusciardi, J.
    [J]. EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2004, 21 : 106 - 106
  • [50] Measurement of the infusion flow rate of a novel non-electrically driven infusion pump in determining the influencing factors on its flow performance
    Cheong, Kar-Hooi
    Doihara, Ryouji
    Furuichi, Noriyuki
    Nakagawa, Masaharu
    Karasawa, Ruriko
    Kato, Yoshihiro
    Kageyama, Kazunori
    Akasaka, Takuro
    Onuma, Yuki
    Kato, Takashi
    [J]. MEASUREMENT, 2023, 218