CLINICAL HIGH-RESOLUTION SKELETAL SINGLE-PHOTON EMISSION TOMOGRAPHY USING A TRIPLE-HEADED GAMMA-CAMERA

被引:6
|
作者
BUSCOMBE, JR
TOWNSEND, CE
KOURIS, K
CLARKE, G
MAHMOOD, S
JARRITT, PH
ELL, PJ
机构
[1] Institute of Nuclear Medicine, Univ College London Medical School, London W1N 8AA, Mortimer Street
来源
BRITISH JOURNAL OF RADIOLOGY | 1993年 / 66卷 / 789期
关键词
D O I
10.1259/0007-1285-66-789-817
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Planar skeletal scintigraphy has become established as a standard diagnostic test performed within the nuclear medicine department. Since the 1970s good quality images have been produced using an Anger gamma camera and Tc-99m-labelled diphosphonates. Single photon emission tomography (SPET) has improved the sensitivity of detection and the ability to localize bony pathology, particularly benign bone disease in the spine. Recently multi-detector gamma cameras dedicated to SPET have become available. One such system, the Toshiba GCA-9300A, has been used to perform routine clinical skeletal SPET in 81 patients. Good quality images have been obtained using an 8 min acquisition in the axial skeleton and a 16 min acquisition protocol in the peripheral skeleton. Multiple sites can be tomographed in the same patient during the same examination using two or more 8 min acquisitions. Such a multi-detector gamma camera offers advantages over the standard single-headed rotating camera for skeletal SPET in terms of both imaging time and image quality. A cost analysis was performed which demonstrated that the additional cost of purchasing such a multidetector gamma camera was less than 30.00 pound per SPET study.
引用
收藏
页码:817 / 822
页数:6
相关论文
共 50 条
  • [1] SINGLE PHOTON-EMISSION TOMOGRAPHY WITH A ROTATING GAMMA-CAMERA
    FAZIO, F
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1980, 53 (628): : 387 - 387
  • [2] SINGLE-PHOTON EMISSION COMPUTED-TOMOGRAPHY (SPECT) OF THE LIVER WITH A ROTATING GAMMA-CAMERA SYSTEM
    BIERSACK, HJ
    KOISCHWITZ, D
    LACKNER, K
    RESKE, SN
    KNOPP, R
    WINKLER, C
    [J]. NUKLEARMEDIZIN, 1981, 20 (05) : 205 - 213
  • [3] AN ITERATIVE STRATEGY FOR DETERMINATION OF THE SOURCE DISTRIBUTION IN SINGLE-PHOTON EMISSION TOMOGRAPHY WITH A ROTATING GAMMA-CAMERA (SPECT)
    LUIG, H
    ESCHNER, W
    BAHRE, M
    VOTH, E
    NOLTE, G
    [J]. NUKLEARMEDIZIN, 1988, 27 (04) : 140 - 146
  • [4] THE INTERFACING OF A GAMMA-CAMERA TO A DEC GAMMA-11 DATA-PROCESSING SYSTEM FOR SINGLE-PHOTON EMISSION TOMOGRAPHY
    CHESSER, R
    GEMMELL, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1982, 27 (03): : 437 - 441
  • [5] QUALITY-CONTROL IN SINGLE PHOTON-EMISSION TOMOGRAPHY USING A ROTATING GAMMA-CAMERA
    KRETSCHKO, J
    NIELAND, P
    [J]. NUKLEARMEDIZINER, 1991, 14 (03): : 113 - 120
  • [6] Imaging of dopamine transporters in rats using high-resolution pinhole single-photon emission tomography
    Jan Booij
    Kora de Bruin
    Jan B. Habraken
    Pieter Voorn
    [J]. European Journal of Nuclear Medicine and Molecular Imaging, 2002, 29 : 1221 - 1224
  • [7] Imaging of dopamine transporters in rats using high-resolution pinhole single-photon emission tomography
    Booij, J
    de Bruin, K
    Habraken, JBA
    Voorn, P
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (09) : 1221 - 1224
  • [8] VERY HIGH-RESOLUTION GAMMA-CAMERA PROTOTYPE
    PANI, R
    SCOPINARO, F
    DEPAOLA, G
    VACCARO, R
    VALLETTA, L
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1992, 19 (08): : 584 - 584
  • [9] A COMPARISON OF BRAIN IMAGING WITH GAMMA-CAMERA, SINGLE-PHOTON EMISSION COMPUTED-TOMOGRAPHY, AND TRANSMISSION COMPUTED-TOMOGRAPHY - CONCISE COMMUNICATION
    WATSON, NE
    COWAN, RJ
    BALL, MR
    MOODY, DM
    LASTER, DW
    MAYNARD, CD
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1980, 21 (06) : 507 - 511
  • [10] GAMMA-CAMERA EMISSION TOMOGRAPHY USING RADIOLABELED ANTIBODIES
    PERKINS, AC
    WHALLEY, DR
    BALLANTYNE, KC
    PIMM, MV