ON HIGH-RESOLUTION TRANSMISSION ELECTRON-MICROSCOPY IN AN UNSTABLE MAGNETIC ENVIRONMENT

被引:4
|
作者
LUZZI, DE [1 ]
机构
[1] UNIV PENN,RES STRUCT MATTER LAB,PHILADELPHIA,PA 19104
来源
关键词
Feedback circuit; Helmholtz coil; HREM;
D O I
10.1002/jemt.1060150411
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Owing to the particular environment of the JEM4000EX HREM at the University of Pennsylvania, it became necessary to control the magnetic environment around the microscope. The unstable magnetic environment included two components, a slowly wandering vertical DC magnetic field and an AC magnetic field. The effects of these fields on the microscope performance were to introduce an uncertainty in the objective lens defocus value over a series of images and to reduce the attainable resolution of the microscope, respectively. A solution is presented in which these fields are stably reduced well within the limits of sensitivity of the JEOL, Ltd. JEM4000EX for highresolution imaging. This solution is based on a feedback loop using a pseudoHelmholtz coil to generate a welldefined vertical magnetic field. Copyright1990 WileyLiss, Inc.