ACCURATE INFRARED TRANSMITTANCE MEASUREMENTS ON OPTICAL FILTERS USING AN FT-IR SPECTROMETER

被引:8
|
作者
COMPTON, DAC [1 ]
DRAB, J [1 ]
BARR, HS [1 ]
机构
[1] HUGHES AIRCRAFT CO,CARLSBAD,CA 92008
来源
APPLIED OPTICS | 1990年 / 29卷 / 19期
关键词
D O I
10.1364/AO.29.002908
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The accurate measurement of transmittance characteristics of optical filters in the IR region of the spectrum is a demanding application for an IR spectrometer. We have demonstrated that the well known advantages of the Fourier tremsform IR spectrometer (FT-IR) can be utilized for this application. Measurements using FT- IR may he made faster and with better SNR than with dispersive instruments. In addition, the measure¬ments have an optical precision that is superior in wavelength accuracy and equal to, or better, in transmittance accuracy. Instrumental parameters and test procedures necessary to obtain reliable and reproducible data are discussed in detail. The spectral characteristics of filters in both the near- and mid-IR regions are presented, covering a total range from 1-50 fim (10, 000-200 wavenumbers). The types of optical filters examined included those with narrow and wide bandpass, and shortpass characteristics. © 1990 Optical Society of America.