Long-Term Follow-Up of Endocardial Pacing Leads Implanted with Extrathoracic Subclavian Venous Approach

被引:0
|
作者
Ohnuki, Masahiro [1 ]
Miyataka, Kazuhiko [1 ]
Nakamura, Takehiko [2 ]
Ohnishi, Yoshinobu [3 ]
Kohnoike, Yoshizumi [4 ]
Tsujimoto, Masayuki [4 ]
Shirayama, Reirou [5 ]
机构
[1] Saiseikai Chuwa Hosp, Dept Cardiol, Abe 323, Sakurai, Nara, Japan
[2] Takita Hosp, Dept Internal Med, Nara, Japan
[3] Tenri City Hosp, Dept Internal Med, Tenri, Nara, Japan
[4] Heisei Mem Hosp, Dept Internal Med, Unnan, Japan
[5] Nara City Hosp, Dept Internal Med, Nara, Japan
关键词
Pacemaker; Endocardial lead; Lead failure; Lead survival;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The extrathoracic subclavian venous approach is a new approach with the aim to exclude the disadvantages of the conventional subclavian venous approach that carries a risk of pneumothorax and may result in lead damage. We investigated long-term survival of the leads implanted by this new approach between May 1995 and June 2005. A total of 482 leads implanted in 309 patients were analyzed. In cases of patient death, meeting criteria for lead failure or difficulty of continuing follow-up for other reasons, the follow-up was terminated at that time. Therefore, the follow-up was complete for 358 leads (74.3%) and the mean follow-up time was 55.0 +/- 32.1 months (range 1-122 months). The overall lead survival rate was 100% at 5 years and 98.3% at 7 years. This finding suggested that this approach might reduce the incidence of lead failure.
引用
收藏
页码:110 / 114
页数:5
相关论文
共 50 条
  • [1] LONG-TERM FOLLOW-UP OF TRANSVENOUS ENDOCARDIAL ATRIAL LEADS
    KRUSE, IM
    RYDEN, L
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (03): : 702 - 702
  • [2] LONG-TERM FOLLOW-UP OF VENTRICULAR ENDOCARDIAL PACING LEADS - COMPLICATIONS, ELECTRICAL PERFORMANCE, AND LONGEVITY OF 561 RIGHT VENTRICULAR LEADS
    KAZAMA, S
    NISHIYAMA, K
    MACHII, M
    TANAKA, K
    AMANO, T
    NOMURA, T
    OHUCHI, M
    KASAHARA, S
    NIE, M
    ISHIHARA, A
    JAPANESE HEART JOURNAL, 1993, 34 (02): : 193 - 200
  • [3] LONG-TERM PERFORMANCE OF ENDOCARDIAL PACING LEADS
    HELGUERA, ME
    MALONEY, JD
    PINSKI, SL
    WOSCOBOINIK, JR
    WILKOFF, BL
    CASTLE, LW
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (01): : 56 - 64
  • [4] Long-term follow-up of unipolar configuration in bipolar mixed and polyurethane pacing leads
    Verlato, R
    Bongiorni, MG
    Mantovan, R
    Levorato, D
    Baccillieri, S
    Pantaleoni, A
    Arena, G
    Soldati, E
    Mariani, M
    PACEMAKER LEADS 1997, 1997, : 423 - 426
  • [5] Long-term follow-up of DDD pacing mode
    Ulman, Mateusz
    Debski, Maciej
    Zabek, Andrzej
    Haberka, Kazimierz
    Lelakowski, Jacek
    Malecka, Barbara
    KARDIOLOGIA POLSKA, 2014, 72 (06) : 519 - 526
  • [6] LONG-TERM FOLLOW UP OF ENDOCARDIAL PACING IN CHILDREN OF 15 KG OR LESS
    TILL, JA
    JONES, S
    SHINEBOURNE, EA
    WARD, DE
    BRITISH HEART JOURNAL, 1989, 61 (05): : 467 - 467
  • [7] Epicardial pacing leads: the fractal design in ELC 54 improves pacing thresholds in a long-term follow-up
    Novak, M
    Janousek, J
    Smola, M
    Psenicka, M
    Vanek, I
    PACEMAKER LEADS 1997, 1997, : 349 - 352
  • [8] Long-term follow-up of biventricular pacing using a totally endocardial approach in patients with end-stage cardiac failure
    Pasquie, J. L.
    Massin, F.
    Macia, J. C.
    Gervasoni, R.
    Bortone, A.
    Cayla, G.
    Grolleau, R.
    Leclercq, F.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2007, 30 : S31 - S33
  • [9] Mid-term follow-up of endocardial biventricular pacing
    Jaïs, P
    Takahashi, A
    Garrigue, S
    Yamane, T
    Hocini, M
    Shah, DC
    Barold, SS
    Deisenhofer, I
    Haïssaguerre, M
    Clémenty, J
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (11): : 1744 - 1747
  • [10] LONG-TERM PERFORMANCE OF EPIMYOCARDIAL PACING LEADS IN ADULTS - COMPARISON WITH ENDOCARDIAL LEADS
    HELGUERA, ME
    MALONEY, JD
    WOSCOBOINIK, JR
    TROHMAN, RG
    MCCARTHY, PM
    MORANT, VA
    WILKOFF, BL
    CASTLE, LW
    PINSKI, SL
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03): : 412 - 417