TEMPERATURE AND HEAT FLOW RATE CALIBRATION PROCESS OF A DIFFERENTIAL SCANNING CALORIMETER

被引:0
|
作者
Reynoso, Pablo E. [1 ]
Danon, Claudio A. [1 ]
机构
[1] Comis Nacl Energia Atom, Gerencia Mat, Ctr Atom Constituyentes, Ave Gral Paz 1499,B1650KNA, Buenos Aires, DF, Argentina
来源
AVANCES EN CIENCIAS E INGENIERIA | 2016年 / 7卷 / 04期
关键词
thermal analysis; differential scanning calorimetry; instrument calibration; thermal cycles;
D O I
暂无
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The differential scanning calorimeter is not an absolute measuring device, so that it is indispensable to calibrate it for a suitable utilization. In this contribution a calibration procedure for which temperatures and heats of fusion accepted in the literature of were considered five pure materials is presented. Six samples of different masses were prepared for each material having different masses and each of them was submitted to heating from room temperature up to 50 degrees C above of the end fusion temperature using various heating rates for each sample. Then, to determine a ratio of fit between the experimental data obtained and accepted values, a method of nonlinear regression was used by a matrix procedure developed in the MATLAB environment. The results showed a correction with a standard deviation of 1,2 degrees C in temperature and 2,15 J/g in heat flow rate.