NONINVASIVE DETECTION OF SUCTION IN AN IMPLANTABLE ROTARY BLOOD PUMP USING NEURAL NETWORKS

被引:16
|
作者
Karantonis, Dean [1 ]
Cloherty, Shaun [2 ]
Lovell, Nigel [2 ]
Mason, David [3 ]
Salamonsen, Robert [4 ]
Ayre, Peter [5 ]
机构
[1] Univ New South Wales, Sch Elect Engn & Telecommun, Sydney, NSW 2052, Australia
[2] Univ New South Wales, Grad Sch Biomed Engn, Sydney, NSW 2052, Australia
[3] Monash Univ, Dept Surg, Melbourne, Vic, Australia
[4] Alfred Hosp, Prahran, Vic 3181, Australia
[5] Ventracor Ltd, Chatswood, NSW 2067, Australia
关键词
Implantable rotary blood pump; suction detection; control strategy; left ventricular assist device;
D O I
10.1142/S1469026808002302
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Granting those heart failure patients who are recipients of an implantable rotary blood pump (iRBP) greater functionality in daily activities is a key long-term strategy currently being pursued by many research groups. A reliable technique for noninvasive detection of the various pumping states, most notably that of ventricular collapse or suction, is an essential component of this strategy. Presented in this study is such a technique, whereby various indicators are derived from the noninvasive pump feedback signals, and a suitable computational methodology developed to classify the pumping states of interest. Clinical telemetry data from ten implant recipients was categorized (with the aid of trans-oesophageal echocardiography) into the normal and suction states. These data are used to develop a pumping state classifier based on an artificial neural network (ANN). Nine indices, derived from the noninvasive impeller speed signal, form the inputs to this ANN classifier. During validation, the resulting ANN classifier achieved a maximum sensitivity of 98.54% (609/618 samples of 5 s in length) and specificity of 99.26% (12,123/12,213 samples) for correct detection of the suction state. The ability to detect the suction state with such a high degree of accuracy provides a critical parameter both for control strategy development, and for clinical care of the implant recipient.
引用
收藏
页码:237 / 247
页数:11
相关论文
共 50 条
  • [21] Rotary Blood Pump Control Strategy for Preventing Left Ventricular Suction
    Wang, Yu
    Koenig, Steven C.
    Slaughter, Mark S.
    Giridharan, Guruprasad A.
    [J]. ASAIO JOURNAL, 2015, 61 (01) : 21 - 30
  • [22] Feedback Control of Rotary Blood Pump for Preventing Left Ventricular Suction
    Son, Jeongeun
    Du, Dongping
    Du, Yuncheng
    [J]. 2019 AMERICAN CONTROL CONFERENCE (ACC), 2019, : 5426 - 5431
  • [23] Fault Identification in a Blood Pump Using Neural Networks
    Kuehn, Jan
    Buglowski, Mateusz
    Stollenwerk, Andre
    Kowalewski, Stefan
    Walter, Marian
    Leonhardt, Steffen
    Petran, Jan
    Kopp, Ruedger
    Rossaint, Rolf
    Janisch, Thorsten
    [J]. WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 2, 2019, 68 (02): : 27 - 32
  • [24] Experimental determination of dynamic characteristics of the VentrAssist implantable rotary blood pump
    Chung, MKH
    Zhang, N
    Tansley, GD
    Qian, Y
    [J]. ARTIFICIAL ORGANS, 2004, 28 (12) : 1089 - 1094
  • [25] Initial in vivo experience of the VentrAssist implantable rotary blood pump in sheep
    van der Meer, AL
    James, NL
    Edwards, GA
    Esmore, DS
    Rosenfeldt, FL
    Begg, JD
    Woodard, JC
    [J]. ARTIFICIAL ORGANS, 2003, 27 (01) : 21 - 26
  • [26] Modeling and control of an implantable rotary blood pump for heart failure patients
    AlOmari, Abdul-Hakeem H.
    Savkin, Andrey V.
    Ayre, Peter J.
    Lim, Einly
    Lovell, Nigel H.
    [J]. 2010 AMERICAN CONTROL CONFERENCE, 2010, : 3574 - 3579
  • [27] Multisensors information fusion with neural networks for noninvasive blood glucose detection
    Wang, W
    Yan, LF
    Liu, BW
    Zhang, H
    [J]. ADVANCES IN NEURAL NETWORKS - ISNN 2005, PT 3, PROCEEDINGS, 2005, 3498 : 760 - 766
  • [28] Automatic system for noninvasive blood pressure determination in rotary pump recipients
    Schima, T
    Boehm, T
    Huber, L
    Schmallegger, H
    Vollkron, M
    Hiesmayr, M
    Noisser, T
    Wieselthaler, 'G
    [J]. ARTIFICIAL ORGANS, 2004, 28 (05) : 451 - 457
  • [29] A Control System for Rotary Blood Pumps Based on Suction Detection
    Ferreira, Antonio
    Boston, J. Robert
    Antaki, James F.
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2009, 56 (03) : 656 - 665
  • [30] Modeling Study of the Failing Heart and its Interaction with an Implantable Rotary Blood Pump
    Ramachandran, Deepa P.
    Luo, Chuan
    Ma, Tony S.
    Clark, John W., Jr.
    [J]. 2011 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2011, : 2403 - 2409