PRECISION OF OFFICE-BASED BLOOD-GLUCOSE METERS IN SCREENING FOR GESTATIONAL DIABETES

被引:27
|
作者
CARR, SR
SLOCUM, J
TEFFT, L
HAYDON, B
CARPENTER, M
机构
[1] Division of Maternal-Fetal Medicine, Women and Infants Hospital, Brown University School of Medicine Providence, RI
关键词
GESTATIONAL DIABETES SCREENING; TEST PRECISION; BLOOD GLUCOSE METERS;
D O I
10.1016/0002-9378(95)91368-8
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
OBJECTIVE: Is the precision of currently available glucose meters adequate for gestational diabetes screening? STUDY DESIGN: We performed a prospective cohort study of 62 gravid women and compared the precision of three glucose meters to laboratory standard technology. RESULTS: The HemoCue coefficient of Variation was sufficiently low in venous whole blood and plasma and capillary whole blood and plasma (3.0%, 1.8%, 2.8%, and 2.3%, respectively) to allow use in gestational diabetes screening. The One Touch II coefficient of variation was sufficiently low in capillary whole blood (2.7%) to allow use in gestational diabetes screening but was too high in Venous whole blood and plasma and in capillary plasma (3.6%, 7.8%, and 5.7%). The ExacTech coefficient of variation was too high in venous or capillary whole blood (5.4% and 8.1%) for use in gestational diabetes screening and does not analyze plasma samples. CONCLUSION: The HemoCue meter in venous and capillary whole blood and plasma and the One Touch II meter in capillary whole blood have coefficients of variation low enough for gestational diabetes screening.
引用
收藏
页码:1267 / 1272
页数:6
相关论文
共 50 条
  • [21] PORTABLE BLOOD-GLUCOSE METERS - TEACHING PATIENTS HOW TO CORRECTLY MONITOR DIABETES
    BURRITT, MF
    HANSON, E
    MUNENE, NE
    ZIMMERMAN, BR
    [J]. POSTGRADUATE MEDICINE, 1991, 89 (04) : 75 - &
  • [22] SCREENING FOR GESTATIONAL DIABETES-MELLITUS - USE AND ACCURACY OF CAPILLARY BLOOD-GLUCOSE MEASURED WITH A REFLECTANCE METER
    YOO, TW
    CHAO, J
    [J]. JOURNAL OF FAMILY PRACTICE, 1989, 29 (01): : 41 - 44
  • [23] Analysis on the Correlation between Fasting Blood-glucose and Gestational Diabetes Mellitus
    Chen, Sufang
    Wang, Chunyan
    Ren, Xiuru
    Zhu, Huifang
    [J]. 2014 2ND INTERNATIONAL CONFERENCE ON SOCIAL SCIENCE AND HEALTH (ICSSH 2014), PT 3, 2014, 57 : 54 - 57
  • [24] The importance of fasting blood glucose in screening for gestational diabetes
    Herrera, Kimberly
    Brustman, Lois
    Foroutan, Janelle
    Scarpelli, Sophia
    Murphy, Elizabeth
    Francis, Antonia
    Rosenn, Barak
    [J]. JOURNAL OF MATERNAL-FETAL & NEONATAL MEDICINE, 2015, 28 (7-8): : 825 - 828
  • [25] Postnatal Blood Glucose Screening in Gestational Diabetes (GDM)
    Jarvis, Sheba
    Feben, Chris
    Roy, Stephanie
    Sheridan, Richard
    Boret, Tony
    Clements, Michael R.
    [J]. DIABETES, 2011, 60 : A640 - A640
  • [26] BLOOD-VOLUME REQUIREMENTS OF BEDSIDE BLOOD-GLUCOSE METERS
    NARDOLILLO, C
    YALE, JF
    [J]. DIABETES, 1993, 42 : A25 - A25
  • [27] Office-based diabetes educators
    Baltz, JC
    [J]. DIABETES EDUCATOR, 2004, 30 (01): : 40 - +
  • [28] ARE REFLECTANCE METERS NECESSARY FOR HOME BLOOD-GLUCOSE MONITORING
    FERGUSON, SD
    HUGHES, IA
    [J]. BRITISH MEDICAL JOURNAL, 1980, 281 (6253): : 1494 - 1494
  • [29] ARE REFLECTANCE METERS NECESSARY FOR HOME BLOOD-GLUCOSE MONITORING
    HOME, PD
    ALBERTI, KGMM
    [J]. BRITISH MEDICAL JOURNAL, 1981, 282 (6258): : 149 - 149
  • [30] ARE REFLECTANCE METERS NECESSARY FOR HOME BLOOD-GLUCOSE MONITORING
    FERGUSON, SD
    PROSSER, R
    [J]. BRITISH MEDICAL JOURNAL, 1980, 281 (6245): : 912 - 912