PRECISION OF OFFICE-BASED BLOOD-GLUCOSE METERS IN SCREENING FOR GESTATIONAL DIABETES

被引:27
|
作者
CARR, SR
SLOCUM, J
TEFFT, L
HAYDON, B
CARPENTER, M
机构
[1] Division of Maternal-Fetal Medicine, Women and Infants Hospital, Brown University School of Medicine Providence, RI
关键词
GESTATIONAL DIABETES SCREENING; TEST PRECISION; BLOOD GLUCOSE METERS;
D O I
10.1016/0002-9378(95)91368-8
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
OBJECTIVE: Is the precision of currently available glucose meters adequate for gestational diabetes screening? STUDY DESIGN: We performed a prospective cohort study of 62 gravid women and compared the precision of three glucose meters to laboratory standard technology. RESULTS: The HemoCue coefficient of Variation was sufficiently low in venous whole blood and plasma and capillary whole blood and plasma (3.0%, 1.8%, 2.8%, and 2.3%, respectively) to allow use in gestational diabetes screening. The One Touch II coefficient of variation was sufficiently low in capillary whole blood (2.7%) to allow use in gestational diabetes screening but was too high in Venous whole blood and plasma and in capillary plasma (3.6%, 7.8%, and 5.7%). The ExacTech coefficient of variation was too high in venous or capillary whole blood (5.4% and 8.1%) for use in gestational diabetes screening and does not analyze plasma samples. CONCLUSION: The HemoCue meter in venous and capillary whole blood and plasma and the One Touch II meter in capillary whole blood have coefficients of variation low enough for gestational diabetes screening.