EFFECT OF FAILED DEFIBRILLATION SHOCKS ON ELECTROGRAM AMPLITUDE IN A NONINTEGRATED TRANSVENOUS DEFIBRILLATION LEAD SYSTEM

被引:3
|
作者
BRADY, PA [1 ]
FRIEDMAN, PA [1 ]
STANTON, MS [1 ]
机构
[1] MAYO CLIN & MAYO FDN,DIV CARDIOVASC DIS,ROCHESTER,MN 55905
来源
AMERICAN JOURNAL OF CARDIOLOGY | 1995年 / 76卷 / 08期
关键词
D O I
10.1016/S0002-9149(99)80159-3
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Concern has been raised regarding the ability of a nonthoracotomy integrated lead system to redetect ventricular fibrillation following failed defibrillation shocks due to diminution in postshock intracardiac electrogram amplitude. Whether such a problem could occur with other lead systems is not known, leading to uncertainty regarding a potential ongoing risk of sudden cardiac death in some patients despite implantable cardioverter-defibrillator therapy. To investigate this problem, we measured the amplitude of 10 consecutive ventricular fibrillation endocardial electrograms immediately before and immediately after failed defibrillation shocks in 15 patients at the time of implantation of a nonintegrated, transvenous, pace/sense/defibrillation lead. Overall, mean electrogram amplitude decreased 21%, from 10.7 +/- 4.6 mV before to 8.5 +/- 4.9 mV immediately after failed defibrillation shocks. The change in electrogram amplitude postshock was directly related to shock energy (r = 0.85, p <0.0005), but. shock waveform had no differential effect. Electrogram amplitude could also increase after failed shocks, particularly following those of low energy. No failures to redetect ventricular fibrillation were found. Thus, intracardiac electrogram amplitude is reduced following failed defibrillation shocks in this nonintegrated lead system, but by an amount less than that previously reported for some integrated lead systems. Our findings reveal that failed low energy defibrillation shocks are likely to result in less diminution in postshock intracardiac electrogram amplitude than high energy shocks, and that the postshock amplitude may even Increase after some failed shocks.
引用
收藏
页码:580 / 584
页数:5
相关论文
共 50 条
  • [1] EFFECT OF FAILED DEFIBRILLATION SHOCKS ON DETECTED VENTRICULAR ENDOCARDIAL ELECTROGRAM AMPLITUDE USING A TRANSVENOUS LEAD SYSTEM
    HALPERIN, BD
    OTT, GY
    HAUPT, DW
    FAIN, ES
    LIGON, DA
    SILKA, MJ
    CIRCULATION, 1993, 88 (04) : 154 - 154
  • [2] Lead system optimization for transvenous defibrillation
    Gold, MR
    Foster, AH
    Shorofsky, SR
    AMERICAN JOURNAL OF CARDIOLOGY, 1997, 80 (09): : 1163 - 1167
  • [3] EFFECT OF FIRST-PHASE POLARITY OF BIPHASIC SHOCKS ON DEFIBRILLATION THRESHOLD WITH A SINGLE TRANSVENOUS LEAD SYSTEM
    STRICKBERGER, SA
    MAN, KC
    DAOUD, E
    NEARY, MP
    HORWOOD, LE
    NIEBAUER, M
    HUMMEL, JD
    MORADY, F
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 25 (07) : 1605 - 1608
  • [4] Transvenous defibrillation lead systems
    Gold, MR
    Shorofsky, SR
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (06) : 570 - 580
  • [5] CHRONIC CHANGES IN DEFIBRILLATION PARAMETERS WITH A TRANSVENOUS LEAD SYSTEM
    TUMMALA, RV
    RIGGIO, DW
    WEISS, D
    FELICIANO, Z
    SHOROFSKY, SR
    GOLD, MR
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, : A112 - A112
  • [6] EFFECT OF SHOCK POLARITY ON VENTRICULAR DEFIBRILLATION THRESHOLD USING A TRANSVENOUS LEAD SYSTEM
    STRICKBERGER, SA
    HUMMEL, JD
    HORWOOD, LE
    JENTZER, J
    DAOUD, E
    NIEBAUER, M
    BAKR, O
    MAN, KC
    WILLIAMSON, BD
    KOU, W
    MORADY, F
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (04) : 1069 - 1072
  • [7] THE EFFECT OF POLARITY REVERSAL ON BIPHASIC DEFIBRILLATION THRESHOLD WITH AN INTEGRATED TRANSVENOUS LEAD SYSTEM
    SHOROFSKY, SR
    FOSTER, AH
    GOLD, MR
    CIRCULATION, 1994, 90 (04) : 228 - 228
  • [8] Improved efficacy of anodal biphasic defibrillation shocks following a failed defibrillation attempt
    Roberts, PR
    Allen, S
    Smith, DC
    Urban, JF
    Euler, DE
    Kallok, MJ
    Morgan, JM
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (12): : 1753 - 1759
  • [9] Chronic rise in monophasic defibrillation thresholds with a transvenous lead system
    Kirk, MM
    Shorofsky, SR
    Khalighi, K
    Kavesh, NG
    Peters, RW
    Gold, MR
    AMERICAN JOURNAL OF CARDIOLOGY, 1997, 79 (04): : 502 - &
  • [10] TRANSVENOUS-SUBCUTANEOUS DEFIBRILLATION LEADS - EFFECT OF TRANSVENOUS ELECTRODE POLARITY ON DEFIBRILLATION THRESHOLD
    BLOCK, M
    HAMMEL, D
    BOCKER, D
    BORGGREFE, M
    BUDDE, T
    ISBRUCH, F
    SCHELD, HH
    BREITHARDT, G
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1994, 5 (11) : 912 - 918