Key technologies and the performance measuring methods in variable included angle plane grating monochromator

被引:3
|
作者
Lu Qi-peng [1 ]
Song Yuan [1 ,2 ]
Gong Xue-peng [1 ]
Ma Lei [1 ]
机构
[1] Chinese Acad Sci, Changchun Inst Opt Fine Mech & Phys, State Key Lab Appl Opt, Changchun 130033, Jilin, Peoples R China
[2] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
来源
CHINESE OPTICS | 2016年 / 9卷 / 02期
基金
中国国家自然科学基金;
关键词
ultra-high resolution; variable-included-angle; soft X-ray; synchrotron radiation; plane grating monochromator;
D O I
10.3788/CO.20160902.0284
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The variable-included-angle plane grating monochromator is broadly used in high performance soft X-ray due to its superior performance, such as high throughput and high resolution. With the development Science, improving and optimizing the resolution of variables-included-angle plane grating monochromator is necessary and urgently. In order to study and explore the ultra-high resolution of variables-included-angle plane grating monochromator in synchrotron radiation, some issues were researched in this work. Firstly, the relationship between working modes of variables-included-angle plane grating monochromator with different c(ff) is emphatically researched. According to the relationship, the high resolution working modes of the monochromator can be selected. Secondly, we studied the effects of high heat load on the optical system by using the simulation software and designed an appropriate cooling system in the optical elements to decrease the effect of heat load on monochromator. The results indicated that the slope error of plane mirror declined from 8. 1 mu rad to 3 mu rad. Lastly, we studied the testing methods of the variables-included-angle grating monochromator with the resolution of already reaching 5 x 10(4). And the measuring accuracy of angel is 0. 026 ''. Those studies will provide some help for designing the monochromator with ultra-high resolution in the third generation synchrotron radiation.
引用
收藏
页码:284 / 295
页数:12
相关论文
共 32 条
  • [1] ANDREAS K F., 1996, FACILITY STRU, V4
  • [2] BILDERBACK D H, 2000, J SYNCHROTRON RAD
  • [3] Synchrotron monochromator heating problem, cryogenic cooling solution
    Carpentier, P
    Rossat, M
    Charrault, P
    Joly, J
    Pirocchi, M
    Ferrer, JL
    Kaïkati, O
    Roth, M
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 456 (03): : 163 - 176
  • [4] DORMIANI M., 1988, NUCL INSTRUMENTS MET
  • [5] GAO S S, 2012, ACTA OPT SINICA, V32
  • [6] GONG X P, 2013, ACTA OPT SINICA, V33, P308
  • [7] KHOUNSARY A M, 1997, SPIE INT SOC OPTICAL
  • [8] LI Y, 2012, ACTA OPT SINICA, V32
  • [9] LU LJ, 1994, NUCL INSTRUM METH A, V339, P604, DOI 10.1016/0168-9002(94)90200-3
  • [10] Rotation-angle-accuracy measurement of scanning mechanism in variable included angle plane grating monochromater
    Lu Q.-P.
    Ma L.
    Peng Z.-Q.
    [J]. Guangxue Jingmi Gongcheng/Optics and Precision Engineering, 2010, 18 (07): : 1548 - 1553