Measurement of the Root Canal Length by Endometers ES-02 and ES-03

被引:0
|
作者
Jukic Krmek, Silvana [1 ]
Mihaljevic, Dragana [2 ]
Simeon, Paris [1 ]
Karlovic, Zoran [1 ]
机构
[1] Univ Zagreb, Sch Dent Med, Dept Endodont & Restorat Dent, Gunduliceva 5, HR-10000 Zagreb, Croatia
[2] Univ Zagreb, Sch Dent Med, Zagreb, Croatia
关键词
Root Canal Preparation; Tooth Apex; Root Canal Therapy; Sodium Hypoclorite; Chelating Agents;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The aim of this study was to compare the accuracy of measurements of endometers ES-02 and ES-03 in dry canals and canals filled with sodium-hypochlorite and ethylenediaminetetraacetic acid (EDTA).For the experiment 31 one-rooted human extracted teeth were used. Working length of the root canal was determined visually using a #15 reamer as a control measurement. After that the teeth were embedded in vial with freshly mixed alginate and the measurements were made with both devices. Two measurements were made in dry canal and canal filled with sodium-hypochlorite and EDTA for each device. The first value was the working length determined according to the sign -1,0 on the scale and display, respectively, which corresponds to the apical constriction. The second value was measured according to the sign 0,0 which corresponds to the external foramen. Results were analyzed statistically by MANOVA test, and as post-hoc test Student-t test was used. The measurements same as control were in 65% for ES-02 device, and 61%, for ES-03 device at point -1,0 in dry canal, although the difference was not statistically significant. Fluids in root canal statistically significant negatively influence on accuracy of measurements but that significance was 10 times greater at point of measurement -1,0 than at 0,0.
引用
收藏
页码:19 / 27
页数:9
相关论文
共 26 条
  • [1] Root canal length measurement in teeth with electrolyte compensation
    Nam, KC
    Kim, SC
    Lee, SJ
    Kim, YJ
    Kim, NG
    Kim, DW
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2002, 40 (02) : 200 - 204
  • [2] Electronic Measurement of the Root Canal Length in Primary Molars
    Santos, Lucineide de Melo
    de Araujo, Juliane Mendes
    Limeira dos Reis, Jose Ivo
    Holanda Tenorio, Maria Dania
    PESQUISA BRASILEIRA EM ODONTOPEDIATRIA E CLINICA INTEGRADA, 2009, 9 (02): : 193 - 197
  • [3] Root canal length measurement in teeth with electrolyte compensation
    K. C. Nam
    S. C. Kim
    S. J. Lee
    Y. J. Kim
    N. G. Kim
    D. W. Kim
    Medical and Biological Engineering and Computing, 2002, 40 : 200 - 204
  • [4] Comparison of four electronic root canal length measurement devices
    Subay, Rustem Kemal
    Kara, Ozlem
    Subay, Melike Ordulu
    ACTA ODONTOLOGICA SCANDINAVICA, 2017, 75 (05) : 325 - 331
  • [5] The fundamental operating principles of electronic root canal length measurement devices
    Nekoofar, M. H.
    Ghandi, M. M.
    Hayes, S. J.
    Dummer, P. M. H.
    INTERNATIONAL ENDODONTIC JOURNAL, 2006, 39 (08) : 595 - 609
  • [6] Multifrequency Impedance Method Based on Neural Network for Root Canal Length Measurement
    Qiao, Xiaoyue
    Zhang, Zheng
    Chen, Xin
    APPLIED SCIENCES-BASEL, 2020, 10 (21): : 1 - 13
  • [7] Influence of instrument size and varying electrical resistance of root canal instruments on accuracy of three electronic root canal length measurement devices
    Aggarwal, V.
    Singla, M.
    Bhasin, S. S.
    INTERNATIONAL ENDODONTIC JOURNAL, 2017, 50 (05) : 506 - 511
  • [8] New software: comparison between three software programs for root canal length measurement
    Pour, D. Goodarzi
    Razmi, H.
    Maralani, S. Jabedar
    Zeighami, S.
    DENTOMAXILLOFACIAL RADIOLOGY, 2008, 37 (04) : 228 - 231
  • [9] REDUCED RADIATION-DOSE DURING RADIOGRAPHIC MEASUREMENT OF ROOT-CANAL LENGTH
    HICKEL, R
    VOSS, A
    DEUTSCHE ZAHNARZTLICHE ZEITSCHRIFT, 1986, 41 (10): : 965 - 967
  • [10] The usefulness of ultrasonography as a dynamic measurement system for visualizing root canal working length: an in vivo study
    Eren, Irem
    Deniz, Yesim
    BMC ORAL HEALTH, 2024, 24 (01):