THE IMPROVEMENT OF THE OUTPUT PERFORMANCE OF A LOW-ENERGY MEDICAL STANDING-WAVE ACCELERATOR

被引:7
|
作者
ZHAO, HB
CARTER, R
HANNAFORD, CD
ZHOU, P
HUGHES, EA
机构
[1] UNIV LANCASTER,DEPT ENGN,LANCASTER LA1 4YR,ENGLAND
[2] BEIJING MED EQUIPMENT INST,BEIJING 100011,PEOPLES R CHINA
[3] UNIV OXFORD,DARESBURY LAB,OXFORD,ENGLAND
来源
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT | 1995年 / 356卷 / 2-3期
关键词
D O I
10.1016/0168-9002(94)01248-2
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An important objective in the operation of low energy medical standing wave linear accelerators for medical X-ray use, is to limit the spread of energies into the low energy region in order to reduce skin dose to a minimum. This article suggests how this might be achieved by adjustment of the timing of the electron beam pulse relative to the rf power pulse. A theoretical model of a tube based on equivalent circuits is examined to support this proposal. In addition, parameters which affect the dose rate are optimized, in particular the beam loading, and the coupling constant between the accelerator tube and the rf power source.
引用
收藏
页码:552 / 557
页数:6
相关论文
共 50 条
  • [1] STANDING-WAVE TRIPERIODIC STRUCTURE FOR A 10 MEV MEDICAL ACCELERATOR
    BENSUSSAN, A
    TRAN, DT
    TRONC, D
    NUCLEAR INSTRUMENTS & METHODS, 1977, 140 (02): : 231 - 235
  • [2] PHERMEX STANDING-WAVE LINEAR ELECTRON-ACCELERATOR
    STARKE, TP
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1983, 30 (02) : 1402 - 1404
  • [3] Design of RF Feed System for Standing-Wave Accelerator Structures
    Neilson, Jeffrey
    Tantawi, Sami
    Dolgashev, Valery
    ADVANCED ACCELERATOR CONCEPTS, 2010, 1299 : 463 - 466
  • [4] ANALYTICAL DETERMINATION OF LONGITUDINAL GEOMETRY OF A STANDING-WAVE LINEAR ACCELERATOR
    CHABERT, A
    VEYRON, B
    NUCLEAR INSTRUMENTS & METHODS, 1965, 33 (01): : 154 - &
  • [5] rf distribution system for a set of standing-wave accelerator structures
    Tantawi, Sami G.
    PHYSICAL REVIEW SPECIAL TOPICS-ACCELERATORS AND BEAMS, 2006, 9 (11):
  • [6] Cold test on C-band standing-wave accelerator
    Kim, S. H.
    Yang, H. R.
    Moon, S. I.
    Jang, J.
    Gil, Y. M.
    Cho, M.
    Namkung, W.
    Park, S. J.
    Oh, J. S.
    2007 IEEE PARTICLE ACCELERATOR CONFERENCE, VOLS 1-11, 2007, : 2067 - +
  • [7] Design of RF feed system and cavities for standing-wave accelerator structure
    Neilson, Jeffrey
    Tantawi, Sami
    Dolgashev, Valery
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 657 (01): : 52 - 54
  • [8] BEAM LOADING IN A HIGH-CURRENT STANDING-WAVE ELECTRON ACCELERATOR
    STAPLETON, RE
    GRITZO, LA
    VENABLE, D
    JOURNAL OF APPLIED PHYSICS, 1970, 41 (01) : 82 - +
  • [9] Energy conversion performance in looped and stirling traveling-wave and standing-wave thermoacoustic engines
    Guo, Lixian
    Zhao, Dan
    Yu, Guoyao
    Xu, Jingyuan
    Su, Yunpeng
    Sun, Dakun
    Zhang, Yue
    APPLIED THERMAL ENGINEERING, 2025, 258
  • [10] STANDING-WAVE FREE-ELECTRON LASER 2-BEAM ACCELERATOR
    SESSLER, AM
    WHITTUM, DH
    WURTELE, JS
    SHARP, WM
    MAKOWSKI, MA
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 306 (03): : 592 - 605