TEMPERATURE-MEASUREMENT FOR SCANNING TUNNEL MICROSCOPE SAMPLES USING A DETACHABLE THERMOCOUPLE

被引:7
|
作者
JOHN, KD
WAN, KJ
YATES, JT
机构
来源
关键词
D O I
10.1116/1.586448
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A detachable thermocouple contact for the measurement of sample temperature in ultrahigh vacuum was constructed and tested. The thermocouple was demonstrated to be accurate to within 1.2 K from 77 to 373 K. The thermocouple was also compared against a reference thermocouple of conventional design up to 1300 K during ohmic heating) and the agreement was within 20 K throughout this temperature range.
引用
收藏
页码:2137 / 2138
页数:2
相关论文
共 50 条
  • [1] IMPROVE THERMOCOUPLE TEMPERATURE-MEASUREMENT
    DANEMAN, HL
    HYDROCARBON PROCESSING, 1992, 71 (02): : 100 - 103
  • [2] ELECTRONIC THERMOCOUPLE PROCESSORS IN TEMPERATURE-MEASUREMENT
    WOOD, RF
    WOOD, GH
    ELECTRONIC ENGINEERING, 1979, 51 (629): : 51 - &
  • [3] FAST TEMPERATURE-MEASUREMENT BY INFRARED LINE SCANNING IN A HYPERSONIC SHOCK TUNNEL
    HENCKELS, A
    MAURER, F
    OLIVIER, H
    GRONIG, H
    EXPERIMENTS IN FLUIDS, 1990, 9 (05) : 298 - 300
  • [4] THERMOCOUPLE FOR CONTINUOUS TEMPERATURE-MEASUREMENT OF LIQUID STEEL (E)
    GRUZIN, VG
    KONDRASHIN, VM
    DANISHEVSKII, SK
    MEASUREMENT TECHNIQUES, 1972, 15 (12) : 1876 - 1877
  • [5] A PRACTICAL THERMOCOUPLE FOR TEMPERATURE-MEASUREMENT IN MICROWAVE-OVENS
    VANDEVOORT, FR
    LAUREANO, M
    SMITH, JP
    RAGHAVAN, GSV
    CANADIAN INSTITUTE OF FOOD SCIENCE AND TECHNOLOGY JOURNAL-JOURNAL DE L INSTITUT CANADIEN DE SCIENCE ET TECHNOLOGIE ALIMENTAIRES, 1987, 20 (04): : 279 - 284
  • [6] CORRECTNESS OF TEMPERATURE-MEASUREMENT IN GAS-PHASE REACTION ZONE, USING THERMOCOUPLE
    KUZNETSOV, VA
    KRESTININ, AV
    KASHIRENINOV, OE
    MANELIS, GB
    ZHURNAL FIZICHESKOI KHIMII, 1975, 49 (08): : 2096 - 2098
  • [7] CUTTING TEMPERATURE-MEASUREMENT BY TOOL - WORKPIECE THERMOCOUPLE WITH COMPENSATION OF COLD JUNCTION TEMPERATURE
    PRZEKOPI.S
    MECHANIK MIESIECZNIK NAUKOWO-TECHNICZNY, 1973, 46 (01): : 4 - 6
  • [8] THERMOCOUPLE FOR PROLONGED CONTINUOUS TEMPERATURE-MEASUREMENT OF CARBON-IRON MELT
    ABRAMOV, VN
    KORNIENKO, EN
    ZAVODSKAYA LABORATORIYA, 1975, 41 (04): : 466 - 466
  • [9] MINIATURE FLEXIBLE THERMOCOUPLE FOR TEMPERATURE-MEASUREMENT UP TO 800 DEGREES C
    FRAKTOVN.AA
    GUSHCHIN, GI
    MASLOV, IP
    PRIBORY I TEKHNIKA EKSPERIMENTA, 1974, (03): : 229 - 231
  • [10] ERRORS IN TEMPERATURE-MEASUREMENT BY THERMOCOUPLE PROBES DURING ULTRASOUND INDUCED HYPERTHERMIA
    HYNYNEN, K
    MARTIN, CJ
    WATMOUGH, DJ
    MALLARD, JR
    BRITISH JOURNAL OF RADIOLOGY, 1983, 56 (672): : 969 - 970