共 50 条
Development and Validation of a Stability-Indicating Liquid Chromatographic Method for Determination of Valsartan and Hydrochlorthiazide Using Quality by Design
被引:9
|作者:
Shakya, Ashok K.
[1
]
机构:
[1] Al Ahliyya Amman Univ, Fac Pharm & Med Sci, POB 263, Amman 19328, Jordan
关键词:
Valsartan;
hydrochlorothiazide;
Central Composite Design;
Design of experiments (DoE);
Analytical Quality by Design (AQbD);
Response surface methodology (RSM);
HPLC;
D O I:
10.13005/ojc/320203
中图分类号:
Q5 [生物化学];
Q7 [分子生物学];
学科分类号:
071010 ;
081704 ;
摘要:
The present paper involves the analytical quality by design (AQbD)-based development of a simple, rapid, accurate and precise stability-indicating method for the estimation of valsartan and hydrochlorothiazide. Optimized mobile phase (v/v/v) was water (containing 0.25 ml/L triethylamine), methanol and acetonitrile (50:38:37, pH adjusted to 3.0 +/- 0.1). Chromatographic separation was achieved on Hypersil (R)-Gold C-18 (150 x 4.6 mm, 5 mu m, Thermo Fisher Scientific, USA), column at 25 +/- 2 degrees C. The method was validated as per the "International Conference on Harmonisation" (ICH) guidelines. Chromatographic run time was 10.0 minutes. The linearity range for valsartan and hydrochlorothiazide were 1.25-64.00 mu g/ml and 0.195-10.00 mu g/ml respectively. The limits of detection (LOD) for valsartan and hydrochlorothiazide were 0.253 and 0.0226 mu g/ml while the limits of quantitations (LOQ) were 0.767 and 0.068 mu g/ml respectively, using 10 mu l sample. Stability studies indicate that the degradation of valsartan was higher during oxidative stress than other stress conditions.
引用
下载
收藏
页码:777 / 788
页数:12
相关论文