A COMPARISON OF DIFFERENT RETENTION INDEX SYSTEMS WITH UNSATURATED AND AROMATIC-HYDROCARBONS IN CAPILLARY GAS-CHROMATOGRAPHY ON PEG 20M

被引:4
|
作者
ORAV, A
KUNINGAS, K
RANG, S
机构
[1] Institute of Chemistry, Estonian Academy of Sciences, Tallinn, EE0026
关键词
GAS CHROMATOGRAPHY; RETENTION INDEXES; ADSORPTION EFFECTS; N-TETRADECENES; N-TETRADECYNES;
D O I
10.1007/BF02272257
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The correction of the reproducibility of retention indices on polar capillary columns with different film thickness, using more polar reference standards instead of n-alkanes, was investigated. It was found that the replacement of n-alkanes by 1-alkenes, 3-alkynes and n-alkylbenzenes as standards significantly improves the reproducibility of retention indices and diminishes their temperature increments for n-alkanes, n-alkynes and arenes on PEG 20M columns.
引用
收藏
页码:411 / 414
页数:4
相关论文