REPRODUCIBILITY OF 24-HOUR BLOOD-PRESSURE PROFILE OBTAINED BY AMBULATORY AUTOMATIC-MEASUREMENT

被引:4
|
作者
HOFLING, B [1 ]
GEHRING, J [1 ]
HEINBUCH, S [1 ]
VONHOYNINGENHUENE, K [1 ]
机构
[1] LVA OBERBAYERN,HOHENRIED KLIN,W-8139 BERNRIED,GERMANY
关键词
D O I
10.1055/s-2008-1063755
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Twelve patients (8 males, 4 females; mean age 53 [43-60] years) who were undergoing rehabilitation treatment after myocardial infarction were studied to ascertain the reproducibility of ambulatory automatic blood pressure measurements, All were in the last stage of a rehabilitation programme (tolerance to ordinary activity; symptom-free exercise at 75 W). Within two weeks 2, 3 or 4 blood-pressure profiles over 24 hours (total of 32 readings) were obtained by automatic measurement and the records and mean values were compared. Single mild stresses, such as gymnastic exercise or visit to the doctor, were identifiable on the records, but did not alter the overall profile or mean values. Normotensives, hypertensives and borderline hypertensives had 24-hour profiles which were nearly identical with regard to the curve >>envelope<<, day-night profile and mean values in the individual patients. Mean values of diastolic and systolic pressures day by day in each patient showed no deviations greater than 5 mm Hg. Therapeutic measures were recognizable by parallel fall of the curve >>envelope<<, as well as by a reduction in mean value. - At least in these selected patients a single 24-hour profile would in principle have sufficed to describe blood pressure behaviour.
引用
收藏
页码:1337 / 1341
页数:5
相关论文
共 50 条
  • [1] THE 24-HOUR AMBULATORY BLOOD-PRESSURE PROFILE WITH VERAPAMIL
    GOULD, BA
    MANN, S
    KIESO, H
    SUBRAMANIAN, VB
    RAFTERY, EB
    [J]. CIRCULATION, 1982, 65 (01) : 22 - 27
  • [2] 24-HOUR AMBULATORY BLOOD-PRESSURE MEASUREMENT IN PREGNANCY
    HALLIGAN, AJ
    OBRIEN, E
    OMALLEY, R
    WALSHE, JJ
    DARLING, MRN
    [J]. BRITISH JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 1993, 100 (03): : 290 - 290
  • [3] THE 24-HOUR AMBULATORY PROFILE OF BLOOD-PRESSURE REDUCTION WITH MUZOLIMINE
    CARRAGETA, O
    SOARES, R
    MARTINS, A
    [J]. CLINICAL NEPHROLOGY, 1983, 19 : S114 - S115
  • [4] 24-HOUR AMBULATORY BLOOD-PRESSURE MEASUREMENT IN A PRIMIGRAVID POPULATION
    HALLIGAN, A
    OBRIEN, E
    OMALLEY, K
    MEE, F
    ATKINS, N
    CONROY, R
    WALSHE, JJ
    DARLING, M
    [J]. JOURNAL OF HYPERTENSION, 1993, 11 (08) : 869 - 873
  • [5] 24-HOUR AMBULATORY MONITORY OF BLOOD-PRESSURE
    MEYERSABELLEK, W
    GOTZEN, R
    [J]. ZEITSCHRIFT FUR KARDIOLOGIE, 1991, 80 : R5 - R5
  • [6] 24-HOUR BLOOD-PRESSURE MEASUREMENT
    ZUMTHEMA, PB
    [J]. MUNCHENER MEDIZINISCHE WOCHENSCHRIFT, 1989, 131 (27): : 519 - 519
  • [7] AUTOMATIC 24-HOUR MONITORING OF BLOOD-PRESSURE
    BOGDANOV, MM
    [J]. KARDIOLOGIYA, 1985, 25 (01) : 90 - 93
  • [8] 24-HOUR AMBULATORY BLOOD-PRESSURE IN SHIFT WORKERS
    CHAU, NP
    MALLION, JM
    DEGAUDEMARIS, R
    RUCHE, E
    SICHE, JP
    PELEN, O
    MATHERN, G
    [J]. CIRCULATION, 1989, 80 (02) : 341 - 347
  • [9] 24-HOUR BLOOD-PRESSURE MONITORING IN AN AMBULATORY SETTING
    BOQUINHAS, JM
    BARRETO, JC
    GASPAR, A
    BOQUINHAS, MH
    SIMOES, J
    [J]. KIDNEY INTERNATIONAL, 1994, 45 (01) : 284 - 284
  • [10] AMBULANT 24-HOUR MEASUREMENT OF BLOOD-PRESSURE
    WEHLING, M
    THEISEN, K
    [J]. DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1990, 115 (51-52) : 1960 - 1962