ANALYTE COMPARISONS ON THE COULTER DACOS AND KODAK EKTACHEM DT-60 CLINICAL-CHEMISTRY ANALYZERS

被引:0
|
作者
JACOBS, RM
LUMSDEN, JH
TAYLOR, JA
GRIFT, E
机构
关键词
D O I
暂无
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Eighteen analytes were compared using the Coulter Dacos and Kodak Ektachem DT-60 chemistry analyzers. All analytes were significantly linearly related. Correlations were excellent for fifteen analytes (r > 0.95); the correlations for sodium, chloride, and total protein were less than 0.95 but greater than 0.90. Several analytes had slopes significantly different from unity which was likely related to methodological differences. Regression equations are presented for the conversion of data generated using the Kodak Ektachem DT-60 to Coulter Dacos values so that, in the University of Guelph clinical pathology labortory, one set of reference intervals serves both instruments. These data are directly applicable only in the laboratory in which they were developed. They should not be utilized directly in other laboratories, but can be used as general guidelines until confirmatory studies are done.
引用
收藏
页码:30 / &
相关论文
共 45 条
  • [1] EVALUATION OF THE KODAK EKTACHEM DT-60 ANALYZER
    ELDERINY, S
    NG, RH
    STATLAND, BE
    CLINICAL CHEMISTRY, 1986, 32 (07) : 1415 - 1415
  • [2] PERFORMANCE OF A TABLE-TOP CLINICAL-CHEMISTRY ANALYZER, THE KODAK-EKTACHEM-DT60 ANALYZER
    GREEN, G
    MEADOWS, K
    STANTON, C
    SCHNIPELSKY, P
    SWINIARSKI, K
    CLINICAL CHEMISTRY, 1984, 30 (06) : 951 - 951
  • [3] A SYSTEMS-APPROACH TO MULTISITE DECENTRALIZED CHEMISTRY TESTING UTILIZING KODAK DT-60 ANALYZERS
    JONES, JB
    SHULSKI, A
    SHARRETTS, S
    CLINICAL CHEMISTRY, 1990, 36 (06) : 1076 - 1076
  • [4] EVALUATION OF THE KODAK EKTACHEM HDL KIT WITH CHOLESTEROL ANALYSIS ON THE DT-60 INSTRUMENT
    WARNICK, GR
    LUM, C
    CLINICAL CHEMISTRY, 1986, 32 (06) : 1098 - 1099
  • [5] KODAK EKTACHEM HDL KIT EVALUATED FOR HDL CHOLESTEROL MEASUREMENTS WITH THE DT-60 INSTRUMENT
    WARNICK, GR
    LUM, C
    KNOPP, RH
    CLINICAL CHEMISTRY, 1986, 32 (10) : 1988 - 1988
  • [6] EVALUATION OF THE KODAK EKTACHEM CLINICAL-CHEMISTRY SLIDES FOR LACTATE
    LASKY, FD
    ARTER, T
    CONNELLY, L
    HASSETT, J
    KRAUZA, C
    WARREN, K
    WRENBERG, A
    CLINICAL CHEMISTRY, 1989, 35 (06) : 1132 - 1132
  • [7] AN EVALUATION OF THE KODAK EKTACHEM CLINICAL-CHEMISTRY SLIDE FOR THEOPHYLLINE
    STLOUIS, PJ
    MACDONALD, M
    GIESBRECHT, E
    SOLDIN, SJ
    THERAPEUTIC DRUG MONITORING, 1989, 11 (01) : 93 - 96
  • [8] THE KODAK EKTACHEM DRY LAYER TECHNOLOGY FOR CLINICAL-CHEMISTRY
    REYNOLDS, KM
    UPSALA JOURNAL OF MEDICAL SCIENCES, 1986, 91 (02) : 143 - 146
  • [9] PERFORMANCE OF KODAK EKTACHEM CLINICAL-CHEMISTRY SLIDES FOR TOTAL CHOLESTEROL
    NOVROS, J
    POWERS, D
    FAME, N
    RAND, R
    CLINICAL CHEMISTRY, 1982, 28 (07) : 1608 - 1608
  • [10] ANALYTICAL PERFORMANCE OF THE KODAK EKTACHEM CLINICAL-CHEMISTRY SLIDES FOR THEOPHYLLINE
    NORTON, GE
    NOVROS, JS
    FRICKEY, PH
    CLINICAL CHEMISTRY, 1986, 32 (06) : 1058 - 1058