DETERMINATION OF CADMIUM AND LEAD IN BLOOD REFERENCE SAMPLES BY SIMULTANEOUS GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY

被引:22
|
作者
DEVAL, A [1 ]
SNEDDON, J [1 ]
机构
[1] MCNEESE STATE UNIV,DEPT CHEM,LAKE CHARLES,LA 70609
关键词
D O I
10.1006/mchj.1995.1071
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Cadmium and lead concentrations at low mu g/liter levels were determined in blood reference samples using simultaneous graphite furnace atomic absorption spectrometry with acceptable accuracy. Use of ammonium dihydrogen phosphate allowed an elevated ashing temperature of 600 degrees C to be used, which allowed the blood matrix to be removed without loss of cadmium and lead. The ability to determine two elements simultaneously offers a reduced analysis time. (C) 1995 Academic Press, Inc.