Monte Carlo dose calculation of GZP6 Co-60 stepping source based on a matrix shift technique

被引:9
|
作者
Toossi, Mohammad Taghi Bahreyni [1 ]
Abdollahi, Malihe [1 ]
Ghorbani, Mahdi [1 ]
机构
[1] Mashhad Univ Med Sci, Fac Med, Med Phys Res Ctr, Med Phys Dept, Mashhad, Iran
关键词
Brachytherapy; Co-60; Stepping source; Monte Carlo simulation; Matrix shift based technique;
D O I
10.1016/j.rpor.2010.11.005
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: As a routine method for stepping source simulation, a Monte Carlo program is run according to the number of steps and then the summation of dose from each run is taken to obtain total dose distribution. This method is time consuming. Aim: As an alternative method, a matrix shift based technique was applied to simulate a stepping source for brachytherapy. Materials and methods: The stepping source of GZP6 brachytherapy unit was simulated. In a matrix shift method, it is assumed that a radiation source is stationary and instead the data matrix is shifted based on the number of steps. In this study, by running MCNPX program for one point and calculation of the dose matrix using the matrix shift method, the isodose curves for the esophageal cancer tumor lengths of 4 and 6 cm were obtained and compared with the isodose curves obtained by running MCNPX programs in each step position separately (15 and 23 steps for esophageal cancer tumor lengths of 4 and 6 cm, respectively). Results: The difference between the two dose matrixes for the stepping and matrix shift methods based on the average dose differences are 3.85x10(-4) Gy and 5.19x10(-4) Gy for treatment length of 4 cm and 6 cm, respectively. Dose differences are insignificant and these two methods are equally valid. Conclusions: The matrix shift method presented in this study can be used for calculation of dose distribution for a brachytherapy stepping source as a quicker tool compared to other routine Monte Carlo based methods. (C) 2010 Greater Poland Cancer Centre, Poland. Published by Elsevier Urban & Partner Sp. z.o.o. All rights reserved.
引用
收藏
页码:10 / 13
页数:4
相关论文
共 50 条
  • [1] A Monte Carlo study on dose distribution validation of GZP6 Co-60 stepping source
    Toossi, Mohammad Taghi Bahreyni
    Abdollahi, Maliheh
    Ghorbani, Mahdi
    [J]. REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2013, 18 (02) : 112 - 116
  • [2] Monte Carlo derivation of AAPM TG-43 dosimetric parameters for GZP6 Co-60 HDR sources
    Tabrizi, Sanaz Hariri
    Asl, Alireza Kamali
    Azma, Zohreh
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02): : 153 - 160
  • [3] Dosimetric study of GZP6 60Co high dose rate brachytherapy source
    Lei, Qin
    Xu, Anjian
    Gou, Chengjun
    Wen, Yumei
    He, Donglin
    Wu, Junxiang
    Hou, Qing
    Wu, Zhangwen
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (04): : 313 - 318
  • [4] Radial dose functions of GZP6 intracavitary brachytherapy 60Co sources: treatment planning system versus Monte Carlo calculations
    Mesbahi, A.
    [J]. IRANIAN JOURNAL OF RADIATION RESEARCH, 2008, 5 (04): : 181 - 186
  • [5] Monte Carlo dosimetric study of the Flexisource Co-60 high dose rate source
    Vijande, Javier
    Granero, Domingo
    Perez-Calatayud, Jose
    Ballester, Facundo
    [J]. JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2012, 4 (01) : 34 - 44
  • [6] In-air calibration of new high dose rate Co-60 brachytherapy sources: results of measurements on a GZP6 brachytherapy afterloading unit
    Mesbahi, Asghar
    Naseri, Alireza
    [J]. REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2008, 13 (02) : 69 - 73
  • [7] Monte Carlo Dosimetric Study of the New Flexisource Co-60 High Dose Rate Source
    Vijande, J.
    Granero, D.
    Perez-Calatayud, J.
    Ballester, F.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3812 - 3812
  • [8] Monte Carlo Dosimetric Study of the GZP 60Co Brachytherapy Source with Stainless Steel Applicator
    Wu, J.
    Wang, P.
    Wang, X.
    Kang, S.
    Li, J.
    Tang, B.
    Orlandini, L. C.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : E564 - E564
  • [9] Dosimetric characterizations of GZP6 60Co high dose rate brachytherapy sources: application of superimposition method
    Toossi, Mohammad Taghi Bahreyni
    Ghorbani, Mahdi
    Mowlavi, Ali Asghar
    Meigooni, Ali Soleimani
    [J]. RADIOLOGY AND ONCOLOGY, 2012, 46 (02) : 170 - 178
  • [10] PERIPHERAL DOSIMETRY BY MONTE CARLO FOR 6MV AND CO-60 BEAMS
    Joosten, A.
    Levi, F.
    Baechler, S.
    Mirimanoff, R. O.
    Bochud, F.
    Moeckli, R.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2009, 185 (09) : 636 - 636