DETERMINATION OF METHOPRENE IN WATER SAMPLES BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY

被引:4
|
作者
ALLEN, CR [1 ]
DICKINSON, CM [1 ]
机构
[1] CTR DIS CONTROL, CTR INFECT DIS, DIV PARASIT DIS, CONTROL TECHNOL BRANCH, ATLANTA, GA 30333 USA
来源
JOURNAL OF LIQUID CHROMATOGRAPHY | 1990年 / 13卷 / 02期
关键词
D O I
10.1080/01483919008049550
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A new high-performance liquid-chromatography method has been developed that can determine low concentrations of methoprene in water samples. The method allows for reliable detection of concentrations between 0.005 μg/mL and 0.5 μg/mL. The detection limit for 5-mL samples is 2.5 ng/mL at 255 nm. The standard curve is linear over the entire concentration range. The method was used to do a stability study on aqueous methoprene samples stored at room temperature and under refrigeration. The study shows that aqueous methoprene samples are stable when stored at 4°C for at least three weeks. Copyright © 1990 by Marcel Dekker, Inc.
引用
收藏
页码:371 / 381
页数:11