Physical dosimetry of volumetric modulated arc therapy (VMAT) using EPID and 2D array for quality assurance

被引:4
|
作者
Mohamed, Ismail E. [1 ]
Ibrahim, Ayman G. [2 ]
Zidan, Hamdy M. [3 ]
El-Bahkiry, Hesham S. [4 ,5 ]
El-sahragti, Adel Y. [4 ]
机构
[1] Ain Shams Univ, Fac Med, Radiat Oncol Dept, Cairo, Egypt
[2] Mansoura Univ Hosp, Oncol & Nucl Med Dept, Mansoura, Egypt
[3] Damietta Univ, Phys Dept, Fac Sci, Dumyat, Egypt
[4] Mansoura Univ, Oncol & Nucl Med Dept, Fac Med, Mansoura, Egypt
[5] Jazan Univ, Fac Appl Med Sci, Diagnost Radiol Dept, Jizan, Saudi Arabia
来源
关键词
VMAT; EPID; 2D array; Gamma index; Control charts; QA;
D O I
10.1016/j.ejrnm.2018.02.003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Outline: To address the correspondence of measured and predicted doses for different malignant tumours utilizing various gamma criteria and QA for confirmation of VMAT with an EPID and 2D array detector. Methods: 24 patients with different malignant tumors were treated by VMAT techniques on Varian IX linear accelerator with 6 MV photon beams. Eclipse treatment planning system (TPS) is used to plan Patient's charts. Gamma Index (GI) variation was compared to the procedure of pre-treatment verification in VMAT plans. Results: The gamma criteria (DD/DTA) of dose difference and distance to agreement for (3%/3 mm), mean +/- SD are gamma( <= 1%) = 99.42% +/- 0.67%, gamma(ma)(x) = 2.11 +/- 0.56 and gamma(avg) = 0.19 +/- 0.05 by EPID, and gamma(%<= 1) = 99.36% +/- 0.53%, gamma(max) = 1.65 +/- 0.45 and gamma(avg) = 0.22 +/- 0.05 by using 2D array detector. Conclusions: Specific QA of VMAT patient (using EPID or 2D array) display great possibility to spare time and to verify individual IMRT fields. 3%/3 mm is the most appropriate of gamma criteria (DD/DTA) for VMAT plans quality assurance. Control charts are a beneficial method for verification assessment for patient specific quality control.
引用
收藏
页码:477 / 484
页数:8
相关论文
共 50 条
  • [1] EPID-BASED QUALITY ASSURANCE FOR VOLUMETRIC MODULATED ARC THERAPY
    Hoffmann, L.
    Jorgensen, M. B. Kyed
    Petersen, J.
    Praestegaard, L. H.
    Hansen, R.
    Muren, L. P.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S82 - S83
  • [2] Evaluation of Angular Dependence Correction for 2D Array Detector Using for Quality Assurance of Volumetric Modulated Arc Therapy
    Karthikeyan, N.
    Ganesh, K. M.
    Vikraman, S.
    Shariff, M. H.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 376 - 376
  • [3] Geometric and Dosimetric Quality Assurance in Volumetric Modulated Arc Therapy (VMAT)
    Qian, J.
    Ma, Y.
    Lee, L.
    Xing, L.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S639 - S639
  • [4] Patient Specific Quality Assurance of Volumetric Modulated Arc Therapy (VMAT): Portal Dosimetry Compared with ArcCheck Phantom
    Utitsarn, K.
    Watthanasarn, T.
    Pimthong, J.
    Krongkietlearts, K.
    Pihusut, C.
    Chaloemchawalit, W.
    Jitmon, J.
    Khuanchana, S.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E826 - E826
  • [5] TOWARDS AUTOMATIC QUALITY ASSURANCE (QA) OF VOLUMETRIC MODULATED ARC THERAPY (VMAT)
    Ruchaud, R.
    Rodrigues, C.
    Gibon, D.
    Massoptier, L.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S154 - S154
  • [6] Evaluation of the sensitivity of two 3D diode array dosimetry systems to setup error for quality assurance (QA) of volumetric-modulated arc therapy (VMAT)
    Li, Guangjun
    Bai, Sen
    Chen, Nianyong
    Henderson, Lansdale
    Wu, Kui
    Xiao, Jianghong
    Zhang, Yingjie
    Jiang, Qingfeng
    Jiang, Xiaoqin
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (05): : 13 - 24
  • [7] Verification of patient specific quality assurance system for volumetric modulated arc therapy (VMAT)
    Amoabeng, Kwame Anokye
    Marthinsen, Anne Beate Langeland
    Hasford, Francis
    Tagoe, Samuel Nii Adu
    Anaafi, Evelyn
    [J]. HEALTH AND TECHNOLOGY, 2022, 12 (04) : 779 - 786
  • [8] Verification of patient specific quality assurance system for volumetric modulated arc therapy (VMAT)
    Kwame Anokye Amoabeng
    Anne Beate Langeland Marthinsen
    Francis Hasford
    Samuel Nii Adu Tagoe
    Evelyn Anaafi
    [J]. Health and Technology, 2022, 12 : 779 - 786
  • [9] Investigation of a 2D Ionization Chamber Array for Quality Assurance of Intensity Modulated Arc Therapy
    Koch, N.
    Ashenafi, M.
    Li, J.
    Vanek, K.
    [J]. MEDICAL PHYSICS, 2008, 35 (06) : 2945 - +
  • [10] Application of a new dosimetry formalism to volumetric modulated arc therapy (VMAT)
    Rosser, Karen E.